FAST ACTING APPLANATION TONOMETERS
R. Stuart Mackay
Donner Laboratory, Medical Physics University of California Berkeley 4, Calif.
Arch Ophthalmol. 1963;70(5):733-734.
 |
 |
| Since this article does not have an abstract, we have provided the first 150 words of the full text PDF and any section headings. |
|
 |
 |
To the Editor:
—The article Fast Acting Applanation Tonometers by Schulz, Hilton, and McEwen (Arch Ophthal 69:717-721, 1963) was quite instructive, but perhaps from experience with the Mackay-Marg tonometer, there are a few useful details that might be added.
The theory of these instruments demands that some sort of transducer be surrounded by a coplanar insensitive annulus, and the condition of coplanarity should be well satisfied. Thus, our experiments employed a stiff support and an extremely sensitive transducer. We felt in our original article, that the ionization converter circuit was not sensitive enough (if used in a rugged configuration). The device they describe shows a change of plunger position of 6µ when tilted from upward to downward. As can be seen from one of our later papers (Mackay; Marg; and Oechsli: Nature 194:687-688 [May 19] 1962), this means that there will only be one orientation of this probe for which
. . . [Full Text PDF of this Article]