An Unexpected Problem With an Inexpensive Air Pump for Vitreous Surgery
Serdar A. Özler, MD;
O uz Algan, MD;
Günay Haznedaro lu, MD
Izmir, Turkey
Arch Ophthalmol. 1993;111(5):583-584.
 |
 |
| Since this article does not have an abstract, we have provided the first 150 words of the full text PDF and any section headings. |
|
 |
 |
To the Editor.
—Fluid-air exchange is an important technique in the repair of complicated retinal detachments. Gross and associates1 described a simple, inexpensive, and pressure-regulated air pump for fluid-air exchange during pars plana vitrectomy. Landers and associates2 designed a similar air pump and found it comparable with the Grieshaber Air System (Grieshaber, Langhorn, Pa) in different laboratory and operating room situations. Unfortunately, with a similar air pump we encountered an unexpected problem that should be taken into consideration by vitreous surgeons, especially those in developing countries.
We used an inexpensive fish-tank air pump (Has Research Inc, Izmir, Turkey). The air pump consisted of a transformer for alternating current, a metal rod that was connected to a rare earth magnet, and a diaphragm (Fig 1). The rare earth magnet vibrated the metal rod back and forth (approximately 50 times per second) as alternating current changed its direction. Thus, movements
. . . [Full Text PDF of this Article]