You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | RSS | Access Rights | Sign In


  Vol. 127 No. 4, April 2009 TABLE OF CONTENTS
  Online Only
 •  Online First Table of
Contents
  Clinical Sciences
 •Online Features
 This Article
 •Abstract
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citation map
 •Citing articles on HighWire
 •Citing articles on Web of Science (27)
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Retinal/ Chorioretinal Disorders
 •Prostheses/ Implants
 •Alert me on articles by topic
 Social Bookmarking
  Add to CiteULike Add to Connotea Add to Delicious Add to Digg Add to Facebook Add to Reddit Add to Technorati Add to Twitter What's this?

Feasibility Study of a Retinal Prosthesis

Spatial Vision With a 16-Electrode Implant

Avi Caspi, PhD; Jessy D. Dorn, PhD; Kelly H. McClure, MS; Mark S. Humayun, MD, PhD; Robert J. Greenberg, MD, PhD; Matthew J. McMahon, PhD

Arch Ophthalmol. 2009;127(4):398-401.

ABSTRACT



Objective  To demonstrate that an epiretinal prosthesis can produce patterned visual perception in patients blinded by photoreceptor degeneration who have no other treatment options.

Methods  A totally blind subject with retinitis pigmentosa had a 16-electrode epiretinal prosthesis implanted. The implant is controlled wirelessly by an external computer or a head-mounted video camera. Spatial vision was assessed by measuring the subject's response to direct stimulation of patterns and by comparing the ability of the subject to identify the orientation of gratings with the system on and off.

Results  In response to stimulation of 2 orthogonal rows of electrodes, the subject drew 2 lines with a mean (SEM) angle of 87.4° (1.8°) between them. With the system on, the subject identified the orientation of the grating target up to a spatial resolution that matches the spacing between the adjacent electrodes. In contrast, with the system off, the subject could not detect or identify the target's orientation.

Conclusion  Synchronized stimulation of different retinal locations with an epiretinal prosthesis implanted long-term can produce spatial vision with an acuity level determined by the distance between the electrodes.

Trial Registration  clinicaltrials.gov Identifier: NCT00279500



INTRODUCTION


 Jump to Section
 •Top
 •Introduction
 •Methods
 •Results
 •Comment
 •Author information
 •References

Electrical stimulation of the retina can restore light perception to subjects blinded by degenerative retinal diseases that damage the photoreceptors, such as retinitis pigmentosa or macular degeneration.1 To generate a patterned image, such an implant uses an array of independently controllable electrodes, each stimulating a different area of the retina to create the perception of light in different locations in the visual field. Although retinal prosthesis research and development have progressed greatly over the years,2-4 owing to the small number of subjects with visual prostheses implanted long-term, it had previously not been shown that the brain can perceive a useful spatial image through stimulation of different retinal locations.

The feasibility of a retinal prosthesis to restore vision is currently under investigation in 5 clinical trials. The 16-electrode epiretinal implant from Second Sight Medical Products, Inc, Sylmar, California, has been implanted long-term in 6 subjects in the United States for more than 5 years, and this is the only study with subjects using the full system in their homes (clinicaltrials.govidentifier: NCT00279500; active, not recruiting). Intelligent Medical Implants GmbH, Bonn, Germany, described 4 subjects with an implanted 49-electrode epiretinal array in a trial designed to last 18 months5 (clinicaltrials.gov identifier: NCT00427180; recruiting). The EPI RET3 25-electrode epiretinal array was implanted for 4 weeks in 6 blind subjects6 (clinical trial: DE/CA21/A07). A subretinal implant composed of a 1550-microphotodiode array and a 4 x 4-electrode array from Retina Implant AG, Reutlingen, Germany, was implanted in 8 subjects for 4 weeks7 (clinicaltrials.gov identifier: NCT00515814; recruiting). Second Sight is conducting another long-term clinical trial of a new-generation permanent implant that consists of 60 electrodes (clinicaltrials.gov identifier: NCT00407602; recruiting).

We report the visual performance of a totally blind subject with an implanted first-generation 16-electrode Second Sight retinal prosthesis1, 8 to determine whether his spatial vision could approach the level expected from the spacing between the electrodes in this early prototype. Subjects with the implant experience visual perception as a result of electrical stimulation and are able to perform simple visual tasks such as locating and counting objects.9 We describe a novel measurement of the visual acuity subserved by the retinal prosthesis and quantify the increase in visual acuity performance obtained by stimulating the retina with higher numbers of electrodes.


METHODS


 Jump to Section
 •Top
 •Introduction
 •Methods
 •Results
 •Comment
 •Author information
 •References

This study was granted an investigational device exemption by the US Food and Drug Administration and was approved by the institutional review board of the University of Southern California, Los Angeles. This research adhered to the tenets of the Declaration of Helsinki. Informed consent was obtained, which explicitly stated the investigational nature of both the system and the surgery.

Measurements were carried out on a single subject who had severe retinitis pigmentosa with no light perception and was aged 55 years on the date of the implantation (the only subject active in the follow-up study who was available for testing). Prior to implantation, the subject's level of vision was determined to be no light perception based on a failure to detect any targets during dark-adapted static and kinetic confrontation perimetry and nonrecordable full-field visual evoked potentials and electroretinograms (flash and flicker) in both eyes. The subject received the implant on June 23, 2004, and the results described here were collected between January and August 2007.

The implant consisted of an electronic case and an intraocular stimulating array. The implanted electronic case, a modified version of the Clarion C-1 (Advanced Bionics, Valencia, California), was surgically implanted into the temporal bone as in the cochlear implant.10 Data and power were transmitted to the implant via a 49-MHz inductive wireless link using an external antenna magnetically aligned over the electronic implant. A 16-wire cable ran from the electronic case along a channel drilled along the skull to the orbit. The cable then passed into the orbit through the sclera to the electrode array on the surface of the retina. The intraocular stimulating array contained 16 platinum electrodes in a 4 x 4 arrangement fixed in a clear silicone rubber platform.

The array was positioned over the macula and attached to the surface of the retina using a retinal tack (Figure 1). The stimulation signal sent to each electrode is independently controlled by an external video processing unit. Real-time video captured by a miniature video camera mounted on the subject's glasses is continuously sampled by the video processing unit and the stimulation current amplitude in each electrode is matched to the brightness at the corresponding area of the scene. The sensitivity of subjects to stimulation of individual electrodes is affected by many factors9; this subject's stimulation threshold for the most sensitive electrode was 16 microcoulombs (µC) and the threshold for the least sensitive electrode was 28 µC. The camera zoom is set so that the field of view approximates the angle subtended by the electrode array on the retina. The system can also be operated in direct stimulation mode, where the stimulation pattern sent to each electrode in the array is determined by an external computer instead of by the acquired video image.


Figure 1
View larger version (42K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 1. Implanted 16-electrode array. A, Fundus image of the implanted 16-electrode array. The 260- and 520-µm diameter electrodes were arranged in a checkerboard pattern with center-to-center spacing of 800 µm. B, A frame captured from the video camera during the grating acuity experiment. The yellow lines mark the visual field mapped to each of the electrodes.


Video was recorded with a video splitter (Video Accessory Corp, Boulder, Colorado) and a digital video recorder (Archos 704; Archos, Igny, France). In the grating acuity experiment, images were presented on a 20-in liquid crystal display monitor (Dell 2007FP; Dell, Inc, Round Rock, Texas).


RESULTS


 Jump to Section
 •Top
 •Introduction
 •Methods
 •Results
 •Comment
 •Author information
 •References

EXPERIMENT 1: ORIENTED PATTERNS

The first experiment tested whether the retinal prosthesis could generate a percept of oriented contours. In each trial, direct stimulation mode (sending stimulation directly to the electrodes from a computer, bypassing the camera) was used to stimulate a single row of 4 electrodes for 1 second followed by a 1-second blank interval and then stimulation of a single column of 4 electrodes for 1 second while the subject maintained fixation straight ahead. Immediately after the second stimulus ended, the subject was instructed to use a bright marker to draw the pattern that he perceived on a board placed at arm's length in front of him. Using the system's head-mounted camera, we recorded video of the subject moving the marker on the board. An off-line single-particle tracking program located the position of the marker in each video frame. Because the camera is fixed to the subject's head, the resulting measurements track the subject's description of the 2-dimensional spatial position of the percepts in head-centered coordinates. Figure 2 shows the stimulation pattern and the results of a single trial. In all of the trials (n = 14), the subject drew 2 lines with a mean (SEM) angle of 87.4° (1.8°) between them. These results clearly demonstrate that the brain can identify spatial forms that are determined by the retinotopic organization of electrical stimulation of electrodes in the stimulating array. Furthermore, this pattern recognition is possible without the benefit of head scanning.


Figure 2
View larger version (27K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 2. Measured visual percept as drawn by the subject after stimulation of a row of electrodes followed by stimulation of a column of electrodes (A), with the schematics of the electrodes stimulated during the experiment (B).


EXPERIMENT 2: VISUAL ACUITY

In a second experiment, we measured the resolution limit of the prosthesis by requiring the subject to report the orientation of high-contrast square-wave gratings. In this experiment, the electrode stimulation pattern was determined in real time using the image acquired by the head-mounted video camera. In each session, high-contrast gratings of different spatial scales (2.77-2.00 logMAR; Snellen equivalent 20/11 777-20/2000) were randomly interleaved. Each grating was presented on a computer monitor for 5 seconds, during which time the subject could scan the image using head movements. The subject was then required to choose 1 of the 4 possible orientations (4 alternative forced choice: horizontal, vertical, diagonal to the right, or diagonal to the left). The detection probability was calculated separately for each spatial scale and a logistic psychometric function was fit to the data with 2 free parameters, maximum performance level and threshold (cutoff) spatial frequency. The results are reported in both logMAR and units scaled to the horizontal field of view of the array. With the system off, the subject's performance was not significantly better than chance for all of the spatial frequencies tested (11 of 40 correct for 1.6 logMAR, 14 of 40 correct for 2.0 logMAR, and 14 of 40 correct for 2.5 logMAR; P > .05 for all of the conditions). With the system on, the subject performed significantly above the chance level of 25% (69 of 108 trials correct; P = 1.4 x 10–17 according to the binomial distribution) for grating spatial frequencies up to the critical sampling frequency (Nyquist limit) of the 4 x 4 electrode array (2.21 logMAR or 2 cycles/array) (Figure 3A). When analyzed separately, performance for all of the 4 orientations was above chance for all of the trials up to 2.21 logMAR (95% for horizontal, 63% for diagonal left, 93% for diagonal right, and 93% for vertical). At the critical sampling frequency, each black and white bar of the grating falls on 1 row of electrodes. The resolution limit was therefore well matched to the spacing of the electrodes and performance dropped to chance levels for finer gratings (Figure 3).


Figure 3
View larger version (33K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 3. The percentage correct for the grating acuity task is well matched to the spatial sampling of the electrode array. A, Standard mapping. The stimulation current sent to each electrode is determined by the corresponding area in the video image. B, Reduced-resolution mapping. The video image is mapped to groups of 4 electrodes that are identically stimulated, creating a 2 x 2 array with half of the sampling frequency. C, Arbitrary mapping. The current level sent to each electrode is mapped to a random pixel in the image. A solid line indicates the chance level (25%) and the dashed lines indicate the 95% confidence interval for 72 trials according to the binomial distribution.


To test whether the spatial resolution provided by the system was determined by the fineness of the spacing of the electrodes, we artificially reduced the resolution of the array by stimulating groups of 4 electrodes together while keeping the field of view constant. This effectively created a 2 x 2 array (4 "pixels" total), reducing the resolution of the system by half and the corresponding sampling frequency to 1 cycle/array. The performance of the subject for the simulated 2 x 2 array is shown in Figure 3B. As expected, the subject's performance with the reduced resolution mapping dropped to chance levels at a spatial frequency twice as low as the original 4 x 4 array. When analyzed separately, performance for all of the 4 orientations with the simulated 2 x 2 array was above chance for all of the trials up to 2.52 logMAR (86% for horizontal, 73% for diagonal left, 68% for diagonal right, and 69% for vertical).

Finally, when the structure in the stimulation pattern was decoupled from the video image entirely by arbitrarily associating an area in the image to an electrode in the array (ie, scrambling the video image to the electrode map), performance decreased to chance for all of the spatial frequencies tested (Figure 3C). Because the field of view is identical for all 3 conditions, a specific video input stimulus will stimulate the retina with equal total current levels in the 3 conditions. The only difference between the 3 conditions is the spatial modulation within the electrode array. Arbitrarily mapping the electrodes to areas of the video image reduces the performance to chance, whereas mapping the retinotopic location of each electrode to the corresponding section of the video image results in visual acuity well matched to the retinal prosthesis electrode spacing.


COMMENT


 Jump to Section
 •Top
 •Introduction
 •Methods
 •Results
 •Comment
 •Author information
 •References

We have demonstrated that stimulation of different retinal locations by a long-term epiretinal prosthesis results in the perception of spatial patterns without the benefit of head scanning. The finding that the resolution limit is well matched to the spacing of electrodes in the stimulating array provides strong evidence that the brain perceives a patterned image and not the average brightness resulting from the overall level of stimulation and further suggests that the development of retinal prosthesis systems with more electrodes should provide higher-spatial-resolution vision to blind subjects.

Although the experimental results reported here are from a single subject with a prototype system implanted, these psychophysical measurements conclusively illustrate the spatial capabilities of an epiretinal prosthesis. As additional subjects receive implants of higher-resolution devices, these results will be confirmed and extended to determine how factors such as the state of disease progression, perceptual learning, and improvements in surgical technique and system design contribute to prosthesis-based visual perception.


AUTHOR INFORMATION


 Jump to Section
 •Top
 •Introduction
 •Methods
 •Results
 •Comment
 •Author information
 •References

Correspondence: Matthew J. McMahon, PhD, Second Sight Medical Products, Inc, 12744 San Fernando Rd, Bldg 3, Sylmar, CA 91342 (mm{at}2-sight.com).

Submitted for Publication: August 29, 2008; final revision received December 23, 2008; accepted December 31, 2008.

Financial Disclosure: Drs Caspi, Dorn, Greenberg, and McMahon and Mr McClure are employees of Second Sight Medical Products, Inc, and Dr Humayun is a consultant for Second Sight Medical Products, Inc. All have stock options. Drs Humayun and Greenberg are also investors in Second Sight Medical Products, Inc.

Funding/Support: This work was supported by grant EY012893 from the National Institutes of Health.

Author Affiliations: Second Sight Medical Products, Inc, Sylmar (Drs Caspi, Dorn, Humayun, Greenberg, and McMahon and Mr McClure) and Doheny Eye Institute, Los Angeles (Dr Humayun), California.


REFERENCES


 Jump to Section
 •Top
 •Introduction
 •Methods
 •Results
 •Comment
 •Author information
 •References

1. Humayun MS, Weiland JD, Fujii GY; et al. Visual perception in a blind subject with a chronic microelectronic retinal prosthesis. Vision Res. 2003;43(24):2573-2581. FULL TEXT | WEB OF SCIENCE | PUBMED
2. Tassiker GE. inventor.Retinal stimulator. US patent 2,760,483. August 28, 1956.
3. Zrenner E. Will retinal implants restore vision? Science. 2002;295(5557):1022-1025. FREE FULL TEXT
4. Wickelgren I. Biomedical engineering: a vision for the blind. Science. 2006;312(5777):1124-1126. FREE FULL TEXT
5. Richard G, Keserue M, Feucht M, Post N, Hornig R. Visual perception after long-term implantation of a retinal implant. Invest Ophthalmol Vis Sci. 2008;49:E-abstract 1786.
6. Walter P, Mokwa W, Messner A, EPI RET3 Study Group. The EPI RET3 wireless intraocular retina implant system: design of the EPI RET3 prospective clinical trial and overview. Invest Ophthalmol Vis Sci. 2008;49:E-abstract 3023.
7. Gekeler F, Sachs H, Szurman P; et al. Surgical procedure for subretinal implants with external connections: the extra-ocular surgery in eight patients. Invest Ophthalmol Vis Sci. 2008;49:E-abstract 4049.
8. de Balthasar C, Patel S, Roy A; et al. Factors affecting perceptual thresholds in epiretinal prostheses. Invest Ophthalmol Vis Sci. 2008;49(6):2303-2314. FREE FULL TEXT
9. Yanai D, Weiland JD, Mahadevappa M, Greenberg RJ, Fine I, Humayun MS. Visual performance using a retinal prosthesis in three subjects with retinitis pigmentosa. Am J Ophthalmol. 2007;143(5):820-827. FULL TEXT | WEB OF SCIENCE | PUBMED
10. Papsin BC, Gordon KA. Cochlear implants for children with severe-to-profound hearing loss. N Engl J Med. 2007;357(23):2380-2387. FULL TEXT | WEB OF SCIENCE | PUBMED


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Delicious Delicious   Add to Digg Digg   Add to Facebook Facebook   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter     What's this?

THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES

Frequency and Amplitude Modulation Have Different Effects on the Percepts Elicited by Retinal Stimulation
Nanduri et al.
IOVS 2012;53:205-214.
ABSTRACT | FULL TEXT  

Blind subjects implanted with the Argus II retinal prosthesis are able to improve performance in a spatial-motor task
Ahuja et al.
Br J Ophthalmol 2011;95:539-543.
ABSTRACT | FULL TEXT  

Stimulation with a Wireless Intraocular Epiretinal Implant Elicits Visual Percepts in Blind Humans
Klauke et al.
IOVS 2011;52:449-455.
ABSTRACT | FULL TEXT  

Simulating prosthetic vision: Optimizing the information content of a limited visual display
van Rheede et al.
J Vis 2010;10:32-32.
ABSTRACT | FULL TEXT  

Selective Activation of Neuronal Targets With Sinusoidal Electric Stimulation
Freeman et al.
J. Neurophysiol. 2010;104:2778-2791.
ABSTRACT | FULL TEXT  





HOME | CURRENT ISSUE | PAST ISSUES | TOPIC COLLECTIONS | CME | PHYSICIAN JOBS | SUBMIT | SUBSCRIBE | HELP
CONDITIONS OF USE | PRIVACY POLICY | CONTACT US | SITE MAP
 
© 2009 American Medical Association. All Rights Reserved.