Laser Suture Lysis Through Thick Blebs Using the Blumenthal Lens

Albert S. Khouri, MD; Tatiana I. Forofonova, MD, PhD; Robert D. Fechtner, MD

Arch Ophthalmol. 2006;124:544.

INTRODUCTION

Laser suture lysis is commonly performed after trabeculectomy and is titrated to control postoperative flow through the filter to a target intraocular pressure.1 Originally, the flat edge of a Zeiss 4-mirror lens was used to compress the conjunctiva and Tenon capsule to view the sutures. The Hoskins lens designed for suture lysis improved the view and allowed eyelid elevation for better exposure. The Ritch suture lysis lens was designed to provide good compression and a reasonably wide field of view that helps in locating sutures.2 Recently, the Blumenthal lens was introduced for suture lysis.

The Blumenthal lens (Volk Optical Inc, Mentor, Ohio) has a smaller, rounded-tip design that allows localized compression of tissues (Figure 1). We found that advantageous in improving suture visualization through thickened conjunctiva and Tenon layers. The Blumenthal lens was particularly useful in patients with encapsulated blebs and in black patients where a thick Tenon capsule limits suture visualization (Figure 2). The improved visibility with this lens allowed suture lysis in encapsulated blebs when treatment was not possible with other lenses (Figure 3 and Figure 4).


Figure 500031
View larger version (88K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 1. Blumenthal suture lysis lens with tip showing.



Figure 500032
View larger version (69K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 2. Encapsulated bleb with poor suture visibility.



Figure 500033
View larger version (41K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 3. View through a Ritch lens (A) and a Blumenthal lens (B). Suture is not visible (A) and barely apparent (B), allowing suture lysis through an encapsulated bleb.



Figure 500034
View larger version (33K):
[in this window]
[in a new window]
[as a PowerPoint slide]
 
Figure 4. Suture visibility through a Ritch lens (A) and a Blumenthal lens (B).


The Blumenthal lens provides a reported image magnification of 2 to 3 times. It has an elevated rounded tip that allows excellent compression even in thick encapsulated blebs. Other suture lysis lenses have a less pointed surface and lower reported magnification (Hoskins, Mendelkorn, and Ritch lenses provide x1.2, x1.32, and x1.0, respectively). The added magnification and rounded tip, however, provide a smaller field of view, making it more challenging to locate some sutures. Other lenses, such as the Ritch lens, provide a flatter contact surface and lower magnification and thus a larger field of view.

In our experience, we have found the Blumenthal lens most effective for suture lysis through a thick Tenon capsule and in encapsulated blebs. This lens may be particularly useful in black patients where early encapsulation makes timely suture lysis more difficult to perform.


AUTHOR INFORMATION
 Jump to Section
 •Top
 •Introduction
 •Author information
 •References

Correspondence: Robert D. Fechtner, MD, Institute of Ophthalmology and Visual Science, University of Medicine and Dentistry of New Jersey/New Jersey Medical School, 90 Bergen St, Suite 6100, Newark, NJ 07103 (fechtner{at}umdnj.edu).

Submitted for Publication: April 8, 2005; final revision received May 6, 2005; accepted May 16, 2005.

Financial Disclosure: None.

Author Affiliations: Institute of Ophthalmology and Visual Science, University of Medicine and Dentistry of New Jersey/New Jersey Medical School, Newark.


REFERENCES
 Jump to Section
 •Top
 •Introduction
 •Author information
 •References

1. Melamed S, Ashkenazi I, Glovinski J, Blumenthal M. Tight scleral flap trabeculectomy with postoperative laser suture lysis. Am J Ophthalmol. 1990;109:303-309. PUBMED
2. Ritch R, Potash SD, Liebmann JM. A new lens for argon laser suture lysis. Ophthalmic Surg. 1994;25:126-127. PUBMED