 |
 |

Magnetically Integrated Microporous Implant
Safety and Efficacy of Secondary Posting
Erika Escalona-Benz, MD;
Matthew S. Benz, MD;
Timothy G. Murray, MD;
Brandy C. Hayden, BS;
Eleut Hernandez;
Scott N. Garonzik;
Nicole L. Cicciarelli
Arch Ophthalmol. 2003;121:1596-1600.
ABSTRACT
 |  |
Objective To evaluate the surgical technique and morbidity related to secondary posting of a previously implanted 12-mm microporous high-density polyethylene implant (MEDPOR) and the enhanced motility associated with magnetic coupling of the prosthesis and the implant in a rabbit model.
Methods Eight New Zealand rabbits underwent primary evisceration surgery with implantation of a 12-mm microporous high-density polyethylene implant. At 6 weeks, a 4 x 6-mm stainless steel, titanium-coated post was secondarily inserted in 6 rabbit eyes using a bilevel incision. Four weeks after the second surgery, 3 rabbits were fitted with a magnet-embedded prosthesis. Motility was measured by evaluating lateral prosthetic excursion during direct observation. At 3, 6, and 12 months the implants and surrounding tissues were harvested for histopathologic examination.
Results Secondary placement of the post within the implant was accomplished without difficulty. No signs of erosion, dehiscence, or extrusion were found after a12-month follow-up in any of the study eyes. Clinical grading documented increased movement of the magnetically coupled prostheses compared with nonmagnetically integrated control eyes.
Conclusion This study establishes the safety of secondary posting and the efficacy of magnetically integrated microporous high-density polyethylene implants in the rabbit model.
Clinical Relevance This technique may offer an alternative to patients with previously implanted microporous high-density polyethylene implants seeking enhanced cosmesis and prosthetic motility.
INTRODUCTION
THE FIRST description of enucleation as a treatment of ocular disease was reported in the 16th century.1 Over the years the surgical technique has been greatly modified and enhanced and the use of orbital implants was introduced to achieve improved cosmesis.2-3
The ideal orbital implant should offer excellent cosmesis and motility while minimizing potential risks such as extrusion, infection, and other complications. The search for the ideal implant has resulted in the development of a variety of materials (homologous, heterologous, and alloplastic) and configurations.3 Porous high-density polyethylene implants have become widely used for volume replacement as spherical orbital implants for the treatment of anophthalmia.4-5
Polyethylene is a stable hydrocarbon polymer that is resistant to biological degradation. Porous polyethylene is made from synthetic, ultrahigh-density polyethylene powder that is easily molded into different shapes. Porous spherical implants possess certain qualities that make them unique: they are nontoxic, nonallergenic, biocompatible, and provide a scaffold for ingrowth of fibrovascular tissue.4, 6-7 These properties offer the theoretical advantages of improved adherence to host tissue, reducing the risk of implant extrusion or migration and secondary infection.8-11 Prosthetic motility may also be improved because of enhanced extraocular muscle attachment to the implant.
Microporous implants can accommodate drilling of a tunnel to insert a coupling peg or post that allows the fitting of an ocular prosthesis.12-13 Enhancement of ocular motility is achieved with the integrated ocular prosthesis.13 However, the integration of the implant with a prosthesis by means of a coupling peg has led to high complication rates.14-15
We previously reported the orbital tolerance of a buried polyethylene microporous implant, fitted with an integrated stainless steel post, and the enhanced motility associated with magnetically coupling an ocular prosthesis with the implant in a rabbit model. In this study, integration of the post and magnetic coupling was successfully accomplished with no complications and good motility was achieved in all animals.16 These results form the basis of the current study. The purpose of the present study is to investigate the safety and efficacy and to document the surgical technique of secondary posting of previously implanted microporous high-density polyethylene implants as well as to evaluate the enhanced motility associated with magnetically coupling a buried microporous high-density polyethylene implant with an ocular prosthesis in a rabbit model.
METHODS
All experiments in this study were conducted under the auspices of the Animal Care and Use Committee of the University of Miami School of Medicine, Miami, Fla. Experiments adhered to the guidelines of the Association for Research in Vision and Ophthalmology Statement for the Use of Animals in Ophthalmic and Vision Research.
Eight New Zealand white rabbits, weighing between 4 and 5 kg, underwent primary evisceration of the right eye with implantation of a 12-mm spherical microporous high-density polyethylene implant (MEDPOR; Porex Corp, College Park, Ga). All animals were anesthetized using intramuscular injections of ketamine hydrochloride, acepromazine, and xylazine hydrochloride. They also received topical proparacaine hydrochloride. After evisceration, the 12-mm spherical microporous high-density polyethylene implant was placed inside the scleral shell and the edges were approximated and sutured with a 5-0 polyglactin 910 suture. The deep and superficial Tenon tissues and conjunctiva were closed in a layered fashion with a 6-0 polyglactin 910 suture. A small ophthalmic conformer was inserted and an occlusive dressing was applied. The rabbits were monitored postoperatively for signs of infection and to ensure proper wound healing.
Six weeks after evisceration, under appropriate anesthesia and monitoring, secondary placement of a stainless steel, titanium-coated post was performed in 6 animals using a bilevel incision (Figure 1). After inserting a lid speculum in the operated on right eye, a 5-mm conjunctival incision was made 3 mm above the midline; a second 5-mm incision was made through superficial and deep Tenon tissues 3 mm below the midline (incisions above and below the center at the horizontal meridian). A third incision was made through the sclera at the center midline and tissues were dissected until an area of about 5 x 5 mm of the 12-mm spherical microporous high-density polyethylene implant was visible. A 4 x 6-mm threaded, stainless steel, titanium-coated post (Figure 2) was secured into the implant and embedded flush with the anterior surface of the implant. The orbital structures were irrigated with 60 mg/mL of gentamicin sulfate. The sclera and Tenon tissues were closed with a 5-0 polyglactin 910 suture and the conjunctiva was closed with a running 7-0 polyglactin 910 suture.
|
|
|
|
Figure 1. Surgical technique. Panel A, The 12-mm microporous high-density polylethylene implant is grasped and held in place with an implant stabilizer. A bilevel incision is made. Panel B, A tunnel is drilled on the midline of the implant. A third incision was made through the sclera at the center midline and tissues were dissected. Panel C, The post is secured and embedded flush with the anterior surface of the 12-mm microporous high-density polylethylene implant. Layered closure of the sclera and Tenon tissues is seen.
|
|
|
|
|
|
|
Figure 2. A medical-grade, 4 x 6-mm threaded, stainless steel, titanium-coated post.
|
|
|
Custom-fitted rabbit ocular prostheses were manufactured integrating two 1-mm circular medical-grade rare earth magnets (Figure 3). A spacing interval of 1 mm was maintained and the magnets were embedded 0.5 mm lateral to the midline (right and left of center at the horizontal meridian). The prostheses were vaulted to achieve a 0.5-mm elevation from the central conjunctiva, thereby eliminating direct central apposition of the conjunctiva and prosthesis. An identically designed prosthesis without incorporation of the magnets was used for the control eyes.
|
|
|
|
Figure 3. Prosthesis showing the 2 circular medical-grade rare earth magnets (arrows) embedded 0.5 mm lateral to the midline with a spacing interval of 1 mm.
|
|
|
Four weeks after the secondary implantation of the posts, 3 rabbits were fitted with a magnet-embedded, porcelain prosthesis and 3 control rabbits were fitted with a prosthesis without incorporation of the magnets. Magnetic field strength (force) was determined using a hanging ball technique at varying distances determined to approximate the clinical relation between the prosthetic and the implant.17
Serial examinations of the eye sockets were performed and clinical photographs were taken. Animals were clinically evaluated for prosthetic motility using the implants with a magnetically integrated prosthesis. Control motility was determined by placement of the nonmagnetic prosthetic. Animals were encouraged into extended field gaze using targeted visual stimuli. Motility was graded, in a masked fashion, as good (<2 mm), fair (1-2 mm), or poor (>2 mm), evaluating lateral prosthetic excursion during direct observation. Measurements were graded by the distance in millimeters of the prosthetic limbus from the evaluated eyelid margin (medial and lateral).
Rabbits with secondary implantation of the post were killed 3, 6, and 12 months after the second surgery. The 2 remaining rabbits were killedone at 3 months and the other at 6 months after the initial evisceration surgery. The implant and surrounding tissues were harvested for histopathologic examination and comparison between study groups.
RESULTS
Secondary insertion of a stainless steel, titanium-coated post within the 12-mm microporous high-density polyethylene implant using the bilevel incision technique was accomplished without difficulty or compromise of the implant in all rabbit eyes. All 12-mm microporous high-density polyethylene implants exhibited fibrovascular ingrowth at the time of the second surgery.
Gross clinical evaluation at monthly intervals during a 12-month follow-up period after the second surgery showed no signs of extrusion of the implant or post displacement, dehiscence of the conjunctiva, or infection (Figure 4). The stainless steel, titanium-coated posts were well tolerated. Orbital histopathologic features of the specimens obtained at 3, 6, and 12 months' follow-up disclosed no evidence of conjunctival erosion, inflammation, or other signs of orbital toxic reactions. Muscle attachment to sclera and orbital fat was not compromised.
|
|
|
|
Figure 4. The implanted eye socket 12 months after magnetic coupling. There is no evidence of conjunctival erosion, infection, inflammation, dehiscence, or extrusion.
|
|
|
Clinical grading of all eyes with a magnetically integrated prosthesis documented fair to good motility. All nonmagnetically integrated control eyes were graded as having poor motility. Although this series documented enhanced motility in all eyes with a magnetically integrated prosthesis, the small sample size precludes meaningful statistical analysis. Magnetic force studies documented a coupling force of 77 N, which clinically represents a moderately strong attraction force.
COMMENT
A variety of prosthetic coupling systems have been investigated throughout the years with the goal of achieving excellent cosmesis, good long-term motility, and minimizing complications.12, 15-16,18-19 Transconjunctival integration or "pegging" of the microporous high-density polyethylene implants improves both the motility and the stability of the prosthetic eye in most cases. However, complication rates have been documented to be as high as 48%.14-15 A recent approach to transconjunctival integration or pegging has used titanium alloy motility coupling posts.18-19 Prosthetic motility has been found to be acceptable and this coupling system offers an efficient surgical option in selected cases.19 As with all transconjunctival implants, complications have been reported in short-term follow-up and longer follow-up is needed to assess long-term complication rates.
Magnetically integrated implants have been studied in the distant past by several investigators, but the appearance of late complications, such as exposure and extrusion, led to the discontinuation of this approach.20-21 More recently, our laboratory demonstrated that advanced magnetic coupling of an integrated microporous high-density polyethylene implant with a buried stainless steel post generates clinically effective force profiles with the magnetic prosthesis.16 The design, materials, and placement of the surgical-grade magnets within the prosthesis played an important role in the successful outcomes and absence of complications.
In the present study, secondary insertion of a stainless steel, titanium-coated post into a previously implanted 12-mm microporous high-density polyethylene implant was performed without technical difficulty. No intraoperative complications were noted in any of the treated animals. The bilevel incision technique was associated with rapid socket healing and rehabilitation without postoperative complications such as epithelial breakdown, exposure, or extrusion. Magnetic coupling of the secondary post implanted microporous high-density polyethylene implant with the magnet-embedded prosthesis generated clinically effective force profiles and enhancement of motility, as we have previously reported with an implant integrated with a buried stainless steel post.16
A unique advantage of this coupling system is that it does not involve transconjunctival integration as do the pegging systems. Magnetic coupling between the prosthesis and the postimplanted microporous high-density polyethylene implant is achieved without interruption of the overlying Tenon and conjunctival tissues. Moreover, the vaulted design of the magnet-embedded prosthesis precludes direct apposition of the prosthesis with the buried post which, therefore, may avoid the risks of conjunctival dehiscence, infection, erosion, and extrusion commonly seen with other coupling systems. Longer follow-up is needed to assess long-term complication rates.
The outcomes are promising and this technique may offer an alternative to patients with previously implanted microporous high-density polyethylene implants seeking enhanced cosmesis and prosthetic motility. Weighing the benefits of improved motility and cosmesis with the small risks of infection and extrusion implicit in the performance of a second surgical procedure, it would seem reasonable to consider this approach. A disadvantage to the use of this 12-mm microporous high-density polyethylene implant with a buried stainless steel, titanium-coated post will be the elimination of magnetic resonance imaging studies for these patients.
CONCLUSIONS
This study establishes the new surgical technique, underscoring the feasibility, safety, and efficacy of secondary posting in the rabbit model. Cosmetic enhancement of motility is documented using this magnetically integrated model. These results provide the basis for further clinical investigation of the potential of this technique in improving long-term motility and cosmesis of ocular prostheses on patients with previously implanted microporous high-density polyethylene implants.
AUTHOR INFORMATION
Corresponding author and reprints: Timothy G. Murray, MD, Bascom Palmer Eye Institute, PO Box 016880, Miami, FL 33101 (e-mail: tmurray{at}med.miami.edu).
Submitted for publication August 15, 2002; final revision received June 18, 2003; accepted July 2, 2003.
This study was supported by a small grant from the Porex Corp.
From the Bascom Palmer Eye Institute, Department of Ophthalmology, University of Miami, Miami, Fla (Drs Escalona-Benz, Benz, and Murray and Mss Hayden and Cicciarelli, and Mr Hernandez); and the SNG Prosthetic Eye Institute, Boca Raton, Fla (Mr Garonzik). Mr Garonzik holds a patent on the integrated implant technology, licensed through Porex Corp, College Park, Ga.
REFERENCES
1. Luce CM. A short history of enucleation. Int Ophthalmol Clin. 1970;10:681-687.
2. Gougelmann HP. The evolution of the ocular motility implant. Int Ophthalmol Clin. 1970;10:689-711.
FULL TEXT
3. Hornblass A, Biesman BS, Eviatar JA. Current techniques of enucleation: a survey of 5,439 intraorbital implants and a review of the literature. Ophthal Plast Reconstr Surg. 1995;11:77-88.
ISI
| PUBMED
4. Karesh JW, Dresner SC. High-density porous polyethylene (MEDPOR) as a successful anophthalmic socket implant. Ophthalmology. 1994;101:1688-1696.
ISI
| PUBMED
5. Karcioglu ZA, al-Mesfer SA, Mullaney PB. Porous polyethylene orbital implant in patients with retinoblastoma. Ophthalmology. 1998;105:1311-1316.
FULL TEXT
|
ISI
| PUBMED
6. Rubin PA, Popham JK, Bilyk JR, Shore JW. Comparison of fibrovascular ingrowth into hydroxyapatite and porous polyethylene orbital implants. Ophthal Plast Reconstr Surg. 1994;10:96-103.
ISI
| PUBMED
7. Goldberg RA, Dresner SC, Braslow RA, Kossovsky N, Legmann A. Animal model of porous polyethylene orbital implants. Ophthal Plast Reconstr Surg. 1994;10:104-109.
ISI
| PUBMED
8. Christmas NJ, Gordon CD, Murray TG, et al. Intraorbital implants after enucleation and their complications: a 10-year review. Arch Ophthalmol. 1998;116:1199-1203.
FREE FULL TEXT
9. Christmas NJ, Van Quill K, Murray TG, et al. Evaluation of efficacy and complications: primary pediatric orbital implants after enucleation. Arch Ophthalmol. 2000;118:503-506.
FREE FULL TEXT
10. Anderson RL, Yen MT, Lucci LM, Caruso RT. The quasi-integrated porous polyethylene orbital implant. Ophthal Plast Reconstr Surg. 2002;18:50-55.
PUBMED
11. Van Acker E, De Potter P. Porous polyethylene (MEDPOR) orbital implant: prospective study of 75 primary implantations [in French]. J Fr Ophtalmol. 2001;24:1067-1073.
PUBMED
12. Uyemura MJ, Farris SR, Popham JK, et al. Drilling of a porous polyethylene orbital implant [ARVO abstract]. Invest Ophthalmol Vis Sci. 1996;3:S617.
13. Guillinta R, Vasani SN, Granet DB, Kikkawa DO. Prosthetic motility in pegged versus unpegged integrated porous orbital implants. Ophthal Plast Reconstr Surg. 2003;19:119-122.
PUBMED
14. Kaltreider SA, Newman SA. Prevention and management of complications associated with the hydroxyapatite implant. Ophthal Plast Reconstr Surg. 1996;12:18-31.
ISI
| PUBMED
15. Lin CJ, Liao SL, Jou JR, Kao SC, Hou PK, Chen MS. Complications of motility peg placement for porous hydroxyapatite orbital implants. Br J Ophthalmol. 2002;86:394-396.
FREE FULL TEXT
16. Murray TG, Cicciarelli NL, Croft BH, Garonzik S, Voigt M, Hernandez E. Design of a magnetically integrated microporous implant. Arch Ophthalmol. 2000;118:1259-1262.
FREE FULL TEXT
17. Joondeph HC, Joondeph BC, Mulcahy T. Comparison of three permanent intraocular magnets. Retina. 1992;12:270-272.
FULL TEXT
|
ISI
| PUBMED
18. Purdy EP. Oculoplastic and orbital applications of porous high-density polyethylene implants. Curr Opin Ophthalmol. 1997;8:57-63.
PUBMED
19. Rubin PA, Fay AM, Remulla HD. Primary placement of a motility coupling post in porous polyethylene orbital implants. Arch Ophthalmol. 2000;118:826-832.
FREE FULL TEXT
20. Ellis OH, Levy OR. A new magnetic orbital implant. Arch Ophthalmol. 1956;56:352-360.
FREE FULL TEXT
21. Young JH. Magnetic intra-ocular implant: new surgery of the implant: the magnetic artificial eye. Br J Ophthalmol. 1954;38:705-718.
PUBMED
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
|