Diode Laser Transscleral Cyclophotocoagulation as a Primary Surgical Treatment for Primary Open-angle Glaucoma
Peter R. Egbert, MD;
Seth Fiadoyor, MD;
Donald L. Budenz, MD;
Patience Dadzie, RN;
Sally Byrd, MD
Arch Ophthalmol. 2001;119:345-350.
ABSTRACT
 |  |
Objectives To evaluate the feasibility of diode laser transscleral cyclophotocoagulation
(TSCPC) as a primary surgical treatment for primary open-angle glaucoma and
to compare 2 laser energy settings used for treatment.
Methods In a prospective clinical trial in Cape Coast and Accra, Ghana, 1 eye
of each of 92 patients with primary open-angle glaucoma was treated by diode
laser TSCPC as a primary surgical treatment. Eyes were randomly assigned to
receive treatment by 20 applications of either 1.5 W applied for 1.5 seconds
or 1.25 W applied for 2.5 seconds.
Results Seventy-nine (86%) of 92 patients completed 3 months of follow-up; follow-up
was 13.2 ± 6.0 months (mean ± SD). Intraocular pressure decreased
in 53 (67%) of the 79 eyes. The drop in intraocular pressure was 20% or more
in 37 eyes (47%) and final intraocular pressure was 22 mm Hg or less in 38
eyes (48%). An atonic pupil was a previously unreported complication that
arose in 27 (28%) of 92 eyes. There were no serious complications of hypotony,
phthisis bulbi, or sympathetic ophthalmia. Visual acuity decreased in 18 (23%)
of 79 eyes treated by TSCPC and in 10 (23%) of 47 fellow eyes treated only
with glaucoma medications. There was no difference in outcomes between the
2 laser energy settings.
Conclusions Diode laser TSCPC is a practical, rapid, well-tolerated procedure that
may provide a modest and variable lowering of intraocular pressure. The treatment,
used with conservative energy levels applied to the eye, seems to have few
serious complications, although a previously unrecognized complication of
atonic pupil needs further evaluation. A moderate variation in laser energy
settings does not influence the results of treatment.
INTRODUCTION
DIODE LASER transscleral cyclophotocoagulation (TSCPC) has been used
successfully for the treatment of refractory glaucoma, including those eyes
in which other surgical treatments have failed.1-5
Transscleral cyclophotocoagulation has not been specifically studied for the
primary surgical treatment of primary open-angle glaucoma (POAG), although
its use has been proposed,6 because of the
fear of vision loss after destructive procedures on the ciliary body.
In the black population of West Africa, glaucoma is the major cause
of irreversible blindness. Long-term medical treatment is not feasible in
most regions, however, because of the long distances patients must travel
for treatment, and because of the high expense and low availability of medications.
Glaucoma is often considered to be a surgical problem in these regions. In
Ghana, we have shown that trabeculectomy with antimetabolites is an effective
treatment for POAG.7-9
Unfortunately, Ghana, like many developing countries, has a shortage of both
ophthalmologists comfortable with glaucoma surgery and resources for eye care,
so that trabeculectomy cannot be offered to every glaucoma patient. Because
a simple, rapid, and inexpensive surgical procedure for glaucoma would be
of great benefit, contact diode laser TSCPC offers attractive attributesthe
procedure is quick, easy to learn, and does not require a complicated, sterile
operating room. In addition, the laser is solid state, reliable, compact,
and portable.
We performed a prospective trial of TSCPC for the primary surgical treatment
of POAG in 2 outpatient clinics in Ghana. Our primary goal was to evaluate
the feasibility of diode laser TSCPC as a primary treatment of POAG in developing
countries. A second goal was to determine, by a prospective randomized trial,
whether a change in energy settings would influence results. Because the optimal
laser settingseither for energy or the ratio of power to duration of
exposureare not known, treatment was randomized between 2 settings.
PATIENTS AND METHODS
After approval of the study by the institutional review board of each
clinic, patients with glaucoma were invited to participate in the study. Inclusion
criteria were being older than 20 years and having a diagnosis of POAG. Patients
were excluded if they had previous glaucoma surgery, including argon laser
trabeculoplasty, cataract extraction, or any other ocular surgery, or if they
had no light perception. Most patients had been diagnosed as having glaucoma
only a few days or weeks before treatment and had been using glaucoma medications
for a short time.
All treatments were performed in 2 outpatient clinics in Ghana between
February and August 1997. The Cape Coast Christian Eye Clinic is located in
the coastal city of Cape Coast and the Emmanuel Eye Clinic is in Accra, the
country's capital. Both clinics have developed a reputation for treating glaucoma
and their patients come from a wide geographical area. Trabeculectomies are
performed on patients with glaucoma when possible, but many patients are managed
with glaucoma medications because trabeculectomies cannot be performed in
sufficient numbers.
Glaucoma was diagnosed based on elevated intraocular pressure (IOP)
and glaucomatous optic disc cupping. Most patients came to the clinic for
examination because of a loss of central visual acuity and were found to have
very advanced glaucoma. Ocular examination included uncorrected visual acuity
and visual acuity with glasses, if available; slitlamp examination; IOP by
Goldmann applanator or Tonopen (Mentor O & O Inc, Norwell, Mass); gonioscopy;
and dilated fundus examination. Visual field examinations were not performed.
After a thorough explanation of the procedure, informed consent was
obtained from all patients. One eye of each patient was chosen for treatment
on clinical grounds (often the eye with worse glaucoma). The other eye received
the best available conventional treatment, including glaucoma medications
or trabeculectomy with antimetabolites. Four surgeons (P.R.E., S.F., D.L.B.,
and S.B.) performed the treatment. An OcuLight SLx diode laser (IRIS Medical
Instruments, Mountain View, Calif) with a handheld G-probe (IRIS Medical Instruments)
was used as described by Kosoko et al.2 After
retrobulbar anesthesia, the anterior edge of the footplate was placed at the
corneoscleral limbus and the probe was pressed firmly on the conjunctiva so
that the laser beam was roughly parallel to the visual axis. The conjunctiva
was kept moist with balanced salt solution. Twenty applications were given
in each eye and spaced evenly over 360°. Care was taken to avoid areas
of intense conjunctival pigmentation because they were subject to conjunctival
burns. In this case, the remaining applications were placed closer to each
other to maintain a total of 20. The probe was cleaned with alcohol and used
for 4 treatments.
We randomly assigned patients to receive 1 of 2 energy settings. After
the retrobulbar anesthesia and just before treatment, a nurse tossed a coin
to determine the settings.
Group 1 received a treatment of 1.5 W applied for 1.5 seconds (2.25
J), and group 2 received 1.25 W applied for 2.5 seconds (3.125 J). With 20
applications, each eye received a total of either 45.0 J or 65.5 J. The power
was not adjusted for the occurrence of "pops." These 2 laser parameters were
selected to have different ratios of power to time of exposure, with a moderate
difference in total energy following a small pilot study of several combinations
of settings. In the pilot study, there was a tendency for eyes treated with
higher energies to have excessive postoperative iritis (P.R.E., unpublished
data, 1997). At the end of the procedure, 4 mg of dexamethasone phosphate
was injected subconjunctivally and 1% atropine solution was applied. The patient
then applied topical atropine twice daily and steroid drops 4 times daily
for at least 3 weeks or until iritis resolved. Any glaucoma medications were
stopped after the treatment but restarted as needed.
The change in IOP was the principal outcome of interest both for establishing
overall efficacy and for comparing group 1 with group 2. However, some patients
had relatively normal preoperative IOP of 22 mm Hg or less because of medications;
in these patients the principal outcome was a reduction of medications. Secondary
outcome measures were changes in visual acuity and complications. Patients
were seen at 1 day, 1 week, 3 weeks, and then every 2 to 3 months after treatment.
Patients who failed to keep appointments were contacted by mail when possible
and, in some cases, by clinic personnel traveling to the patient's residence.
Each examination included review of symptoms, number of glaucoma medications,
visual acuity, IOP, and slitlamp examination. The examiners were not masked
to the treatment.
Retreatments were done at the discretion of the follow-up ophthalmologist
if the first treatment failed to decrease the IOP. Retreatment recapitulated
the original laser parameters but was limited to 15 applications over 270°.
Patients were not included in the outcome analysis of IOP unless they
completed at least 3 months of follow-up. We found that IOP varied during
the first 2 months after treatment because of iritis and the use of atropine
and steroids, but by 3 months these influences were absent. If the treated
eye had another glaucoma operation or a cataract extraction, the last preoperative
IOP was recorded and follow-up was terminated.
An increase or decrease in visual acuity was considered to be an increase
or decrease of 2 or more lines on the Snellen chart or 1 category of visual
acuity if a patient was unable to read the eye chart.
2 Tests were used for the categorical variables, and
a nonpaired 2-sided t test for continuous variables.
RESULTS
Ninety-two eyes of 92 patients fit the criteria for inclusion in the
study and were randomly assigned to treatment group 1 (n = 47) or group 2
(n = 45). Table 1 presents the
preoperative characteristics of the 92 enrolled study subjects. All patients
were black and from West Africa.
|
|
|
|
Table 1. Preoperative Data for All Patients
|
|
|
Complete follow-up examinations at 3 months or more were achieved in
79 (86%) of the 92 patients (40 [85%] of 47 patients in group 1, and 39 [86%]
of 45 patients in group 2; P = .83). Those patients
who did not complete the 3-month follow-up compared with those who did were
more often men (11 [85%] of 13 vs 45 [57%] of 79; P
= .06) and older than 50 years (13 [100%] of 13 vs 64 [81%] of 79; P = .09). There was no difference in preoperative IOP between patients
who completed the follow-up and those who did not (P
= .42) or between the clinics that did the treatment (P = .85).
CHANGE IN IOP
The results refer to the 79 treated eyes of 79 patients who completed
at least 3 months of follow-up (Table 2). The follow-up lasted 13.2 ± 6.0 months (mean ±
SD) (range, 3-26 months). Figure 1
shows IOP before treatment and at the last follow-up examination. Intraocular
pressure decreased in 53 (67%) of the eyes; the drop in IOP was 20% or more
in 37 eyes (47%) and 30% or more in 24 eyes (30%). No eye developed an IOP
lower than 10 mm Hg. The mean ± SD pretreatment IOP was 29.0 ±
8.9 mm Hg (range, 17-66 mm Hg), and at the last examination after treatment
IOP decreased to 25.7 ± 10.3 mm Hg (mean ± SD)(range, 10-66
mm Hg) (P = .02). The IOP varied with time in individual
eyes, going both up and down from 1 visit to another. One explanation for
this variation was the poor compliance with taking medications.
|
|
|
|
Table 2. Results After Treatment for Patients Completing at Least 3
Months of Follow-up
|
|
|
|
|
|
|
Intraocular pressure (IOP) in the treated eye before and after diode
laser transscleral cyclophotocoagulation (TSCPC) for all 79 patients who completed
at least 3 months of follow-up.
|
|
|
Intraocular pressure decreased more in eyes with a high pretreatment
IOP than in those with a low pretreatment IOP (regression coefficient, 0.72; P<.001). We also looked at clinically relevant subsets
of the total treated group: eyes with pretreatment IOP greater than 22 mm
Hg and eyes with a pretreatment IOP of 22 mm Hg or less, which were treated
in an attempt to eliminate glaucoma medications (Table 2). A 20% decrease in IOP occurred in 32 (54%) of 59 eyes
in the former group, but in only 3 (15%) of 20 eyes in the latter (Table 2).
Two eyes had an increase in IOP of more than 20 mm Hg that we cannot
fully explain, although both patients had failed to continue taking glaucoma
medications. One patient was a 31-year-old man whose fellow eye had no light
perception and who had an IOP of 47 mm Hg without medications. The treated
eye went from an IOP of 24 mm Hg before treatment with medications to 54 mm
Hg after treatment without medications. The second patient was an 85-year-old
woman whose fellow eye had a trabeculectomy with an IOP of 14 mm Hg. The treated
eye had a pretreatment IOP of 25 mm Hg while taking medications and 66 mm
Hg after treatment without taking medications.
In 32 patients, the fellow eye (the eye not receiving TSCPC) had a trabeculectomy;
in 47 patients the fellow eye was treated with medications only. In these
47 patients, a comparison was made between the eye treated with TSCPC and
the fellow eye treated with medications only (Table 3). The fellow eyes had a small mean increase in IOP.
|
|
|
|
Table 3. Comparison of Eyes Treated by Diode Laser TSCPC With the Fellow
Eye in the Same Patients* (n = 47)
|
|
|
There was no significant difference between the effect of the 2 laser
settings on IOP (Table 4). At
the last examination after treatment, neither the mean IOP, the change in
IOP from baseline, nor the percentage of eyes having a 20% reduction in IOP
was significantly different between the 2 groups.
|
|
|
|
Table 4. Comparison of Treatment Results for Group 1 and Group 2
|
|
|
Several other variables were examined as possible predictors for the
success of treatment. Only pretreatment IOP was a positive predictor as described
above. Neither age, sex, number of "pops," treating clinic, nor postoperative
atonic pupil had a significant effect on the decrease in IOP (P>.25 for each when used as a single predictor in a standard linear
model test).
RETREATMENTS
Only 16 (20%) of 79 eyes were re-treated (7
in group 1 and 9 in group 2). Fourteen of those eyes had 1 retreatment and
2 had 2 retreatments (Table 5).
Using a 20% decrease in IOP as a measure of success, none of the retreated
eyes was a success before retreatment, but 9 (57%) of 16 were a success after
retreatment. (By this measure, 28 [44%] of 63 eyes not re-treated were a success.)
|
|
|
|
Table 5. Results of Retreatment
|
|
|
CHANGES IN VISUAL ACUITY
Visual acuity at the last follow-up examination was compared with visual
acuity before treatment. Eighteen (23%) of 79 eyes had a decrease in visual
acuity, 55 (70%) had no change, and 5 (6%) had an increase. In 1 eye the visual
acuity after treatment was not recorded. Two eyes progressed to no light perception
(both were limited to hand motions before treatment). In the subset of 19
eyes that had relatively good pretreatment visual acuity (20/60 or better),
only 1 (5%) had a decrease in visual acuity. For comparison, a change in visual
acuity was also tabulated for fellow eyes in patients who did not have a trabeculectomy.
One fellow eye progressed to no light perception. There was no significant
difference in visual acuity change between the fellow eyes and the eyes receiving
TSCPC (Table 3).
MEDICATIONS
Eleven (14%) of 79 eyes had a final IOP of 22 mm Hg or less without
medications. In the 20 eyes with an IOP of 22 mm Hg or less before treatment
that were treated in an attempt to eliminate medications, only 2 eyes (10%)
were receiving no medications at the last examination while maintaining IOP
at less than 22 mm Hg.
COMPLICATIONS
Complications were studied in all 92 treated
eyes. Acceptance of the procedure was goodmost patients liked the idea
of a laser and found the procedure less frightening than an operation, such
as trabeculectomy. Most patients experienced mild to moderate pain for a few
days, but none complained of severe pain. Mild iritis was always seen the
day after treatment and severe iritis occurred in 2 eyes, which resolved with
topical steroids within 6 to 8 weeks. Transient conjunctival burns were common.
A hyphema occurred during treatment in 3 eyes but resolved in each case by
the next day. There were no cases of hypotony, phthisis bulbi, or sympathetic
ophthalmia during the follow-up period.
An atonic pupil was noted in 27 (29%) of the 92 eyes (13 from group
1 and 14 from group 2) and may have been present more often because we did
not prospectively look for this characteristic. The change ranged from a mild
irregularity of the pupil, to a segmental lack of pupil constriction, to a
totally nonreactive pupil. When a pupil was distorted toward the limbus, peripheral
anterior synechiae were occasionally seen in this area. However, in other
eyes in which the pupil was distorted, the iris did not react to light in
1 segment and there were no peripheral anterior synechiae. Segmental atony
was most often in the superior iris. Iris transillumination defects were not
seen.
COMMENT
In this prospective study of diode laser TSCPC for the treatment of
POAG in eyes that had not had any previous surgery, we set out to determine
if TSCPC could play a role in the treatment of POAG in developing countries.
The treatment was, as hoped, quick and relatively simple to apply. The laser
was reliable even in face of the tropical heat and humidity of Ghana. Patient
acceptance of the treatment was excellent and patients found the procedure
less frightening than trabeculectomy. The procedure seems safe in that no
major complications were recognized, but the effect of TSCPC on IOP was unpredictable.
The extent of IOP reduction after treatment was modest and varied from
eye to eye. With a mean follow-up of 13.2 months, IOP was reduced 20% or more
in 47% of the eyes, and final IOP was 22 mm Hg or less in 48%. Yet, the IOP
in other eyes remained unchanged or increased. We cannot explain this variation
in full, although a reasonable contributing factor is the erratic use of glaucoma
medications. Preoperative IOP was a significant predictor of success: IOP
was reduced more frequently in eyes with high than with low initial IOP. A
20% IOP reduction occurred in 57% of eyes with a pretreatment IOP greater
then 22 mm Hg and in only 15% of eyes with a pretreatment IOP of 22 mm Hg
or less. One goal of this study was to determine if patients with relatively
normal IOP could stop taking glaucoma medications after treatment. We found
that TSCPC generally does not allow discontinuation of medications in this
group. Just 10% of patients with a pretreatment IOP of 22 mm Hg while taking
medications were able to stop medications and still maintain an IOP of 22
mm Hg or less.
It is important to recognize that retreatments with diode laser TSCPC
have been shown to be effective in other studies and have been used in as
many as 49%3 to 62%10
of eyes. Spencer and Vernon11 have advocated
considering TSCPC as a course of treatment in which multiple sessions are
routine. We retreated only 20% of eyes in our study and found that 57% of
these had a final reduction in IOP of at least 20%. If more eyes that had
not responded to the initial treatment had been re-treated, our final success
rate probably would have been higher. It may be that repeated treatments are
necessary to achieve optimal results.
We randomized treatment between 2 energy levels, with different ratios
of power and duration of application, because the most appropriate laser energy
settings are not known. There was no difference in outcomes between the 2
settings. Krott et al12 also failed to find
a correlation between energy applied during TSCPC and outcome. Within certain
limits, the laser energy settings may not be as crucial to results as other
undetermined variables, such as probe position and angle, or variation in
the anatomy of individual eyes. Bloom et al3
have recommended transillumination to identify the position of the pars plicata
of the ciliary body, but it has not been shown whether this technique improves
results.
We are aware of no other studies that have specifically looked at diode
laser TSCPC as the primary surgical treatment for POAG. There are a few retrospective
studies of TSCPC for refractory glaucoma of various types that have included
results for POAG; these studies report variable results with success rates
for POAG that are usually higher than we found in our study.3, 10-11,13-14
Aside from being retrospective, these studies differed from ours in several
other key waysthe patients were not African, the eyes often were previously
operated on, the treatment energies were usually higher, and retreatments
were more frequent.
In an earlier study, using the same population in Ghana, we examined
trabeculectomy with intraoperative antimetabolites as a treatment of POAG.7-8 Intraocular pressure was reduced to
less than 21 mm Hg in 73% to 93% of eyes with a mean follow-up of 10 months.
Diode laser TSCPC, as applied in the current study, does not match those results.
In terms of complications, TSCPC causes more immediate iritis than trabeculectomy,
but avoids the occasional flat chamber. Late follow-up of trabeculectomy with
antimetabolites in the same population showed a 2% incidence of hypotony9 and no endophthalmitis. We do not have data to compare
cataract formation or vision change between the 2 procedures.
We used conservative amounts of laser energy in the treatment because
there is little information on the complications associated with diode laser
TSCPC in black patients or in POAG. Both the energy per application and the
number of applications were at or below the lower range given in most other
treatment studies. We thought it was important to minimize complications in
patients who had not had other surgery and potentially had other forms of
treatment available for their eyes. Serious complications did not occur; there
were no cases of hypotony (the lowest IOP was 10 mm Hg), phthisis bulbi, or
sympathetic ophthalmia.
Some eyes lost visionvisual acuity decreased in 23% of patients.
There is a strong concern that TSCPC can lead to loss of central vision from
cystoid macular edema or other causes. On the other hand, we have noticed
in our clinical practice in Ghana that patients with advanced glaucoma often
lose vision rapidly as a consequence of the natural progression of the disease.
After diode laser TSCPC, it is difficult to ascertain whether loss of vision
results from treatment or from the natural progression of glaucoma. We attempted
to answer the question by looking at the data in 2 ways. First, the treated
eyes that had relatively good visual acuity (20/60 or better) before treatment
were studiedthese eyes presumably had less risk for rapid loss of visual
acuity from glaucoma than eyes with worse pretreatment visual acuity.
Only
1 (5%) of 19 of these eyes lost visual acuity after TSCPC. Second, we looked
at the vision change in the fellow eyes. The fellow eyes that had treatment
by medications only, with no TSCPC, lost vision at the same rate as the eyes
treated by TSCPC. Precise conclusions cannot be drawn from comparisons of
the treated and untreated eyes because treatment was not randomly allocated
between the eyes. However, the vision loss in fellow eyes is consistent with
the idea that many of the study patients had advanced glaucoma and were at
risk to lose vision from the natural progression of glaucoma. Some may also
have had maturation of cataracts.
An unexpected complication was the atonic pupil that was observed in
29% of treated eyes. Pupil abnormalities have not been reported before except
for one mention of a "localized iris shrinkage with pupillary distortion"
and "atrophy of the iris root" by Mueller et al.13
Atonic pupils may not have been recognized before because the vast majority
of eyes that have been treated with diode laser TSCPC have either had neovascular
glaucoma, which often causes poorly reactive pupils, or have had previous
intraocular operations that distorted the iris. Two cases of neurotrophic
corneal defects have been seen following diode laser TSCPC, which suggests
that corneal sensory nerves can be damaged.15
Perhaps the parasympathetic nerve supply to the iris sphincter was damaged
in our cases. It will be important to monitor patients' symptoms and any loss
of accommodation to evaluate the clinical significance of this finding.
This study has limitations. Because all of the patients were black Africans
and most had very advanced glaucoma, our results may not apply to other populations.
Only 79 (78%) of 92 treated patients completed the 3-month follow-up necessary
to evaluate treatment results, and more patients failed to return for later
examinations. Although this is a good follow-up rate for studies in developing
countries, the patients who did not return could have had different results
than those who completed the follow-up. Also, longer follow-up could reveal
loss of effect or more complications. The use of glaucoma medications was
not constant, which makes it difficult to gauge the effect of TSCPC alone
on IOP.
In conclusion, the role of diode laser TSCPC in the management of POAG
in developing countries warrants further study. It fulfilled some of its expectations
as a practical, rapid, and easy-to-learn primary surgical treatment for POAG.
The treatment, used with conservative energy applied to the eye, seems to
be free from serious complications, although a previously unrecognized complication
of atonic pupil needs further evaluation. There was no difference in the outcome
of the 2 laser energy settings used.
AUTHOR INFORMATION
Accepted for publication August 9, 2000.
This study was supported in part by the Elsie B. Ballantyne Fund, Department
of Ophthalmology, Stanford University. IRIS Medical Instruments supplied the
laser used in the study.
We are grateful for the careful patient care given by K. A. Darko, MD,
and Margaret Appiah, EN, at the Emmanuel Eye Clinic. We thank Bradley Efron,
PhD, Departments of Statistics, and Health Research and Policy, Stanford University,
for assistance with statistical analysis.
Corresponding author and reprints: Peter R. Egbert, MD, Stanford
University Medical Center, Room A 157, Stanford, CA 94305-5308 (e-mail: egbert{at}stanford.edu).
From the Department of Ophthalmology, Stanford University, Stanford,
Calif (Drs Egbert and Byrd); Cape Coast Christian Eye Clinic, Cape Coast,
Ghana (Dr Fiadoyor and Ms Dadzie); and the Department of Ophthalmology, Bascom
Palmer Eye Institute, University of Miami, Miami, Fla (Dr Budenz). The authors
have no financial interest in and no paid consulting arrangement with IRIS
Medical Instruments, Mountain View, Calif.
REFERENCES
 |  |
1. Gaasterland DE, Pollack IV. Initial experience with a new method of laser trans-scleral cyclophotocoagulation
for ciliary ablation in severe glaucoma. Trans Am Ophthalmol Soc. 1992;90:225-243.
PUBMED
2. Kosoko O, Gaasterland DE, Pollack IP, Enger CL the Diode Laser Ciliary Ablation Study Group. Long-term outcome of initial ciliary ablation with contact diode laser
trans-scleral cyclophotocoagulation for severe glaucoma. Ophthalmology. 1996;103:1294-1302.
ISI
| PUBMED
3. Bloom PA, Tsai JC, Sharma K, et al. "Cyclodiode": trans-scleral diode laser cyclophotocoagulation in the
treatment of advanced refractory glaucoma. Ophthalmology. 1997;104:1508-1519.
ISI
| PUBMED
4. Youn J, Cox TA, Herndon LW, Allingham RR, Shields MB. A clinical comparison of trans-scleral cyclophotocoagulation with neodymium:YAG
and semiconductor diode lasers. Am J Ophthalmol. 1998;126:640-647.
FULL TEXT
|
ISI
| PUBMED
5. Hawkins TA, Stewart WC. One-year results of semiconductor trans-scleral cyclophotocoagulation
in patients with glaucoma. Arch Ophthalmol. 1993;111:488-491.
FREE FULL TEXT
6. Higginbotham EJ. Reaffirming the role of the laser in glaucoma management [editorial]. Arch Ophthalmol. 1999;117:1075-1076.
FREE FULL TEXT
7. Egbert PR, Williams AS, Singh KS, Dadzie P, Egbert TB. A prospective trial of intraoperative fluorouracil during trabeculectomy
in a black population. Am J Ophthalmol. 1993;116:612-616.
ISI
| PUBMED
8. Singh KS, Egbert PR, Byrd S, et al. Trabeculectomy with intraoperative 5-fluorouracil vs mitomycin C. Am J Ophthalmol. 1997;123:48-53.
ISI
| PUBMED
9. Singh KS, Byrd S, Egbert PR, Budenz D. Risk of hypotony after primary trabeculectomy with antifibrotic agents
in a black West African population. J Glaucoma. 1998;7:82-85.
ISI
| PUBMED
10. Threlkeld AB, Johnson MH. Contact trans-scleral diode cyclophotocoagulation for refractory glaucoma. J Glaucoma. 1999;8:3-7.
ISI
| PUBMED
11. Spencer AF, Vernon SA. "Cyclodiode": results of a standard protocol. Br J Ophthalmol. 1999;83:311-316.
FREE FULL TEXT
12. Krott R, Diestelhost M, Zollweg M, Krieglstein GK. Dose-response relationship of trans-scleral contact cyclophotocoagulation. Ophthalmologe. 1997;94:273-276.
FULL TEXT
|
ISI
| PUBMED
13. Mueller FO, Mendis U, Jusudasan CAN. Trans-scleral cyclophotocoagulation of uncontrolled glaucoma using
a diode laser and glaucoma probe: follow-up results at up to sixteen months. Ophthalmic Pract. 1997;15:230-235.
14. Werner A, Vick H, Guthoff R. Cyclophotocoagulation with the diode laser. Ophthalmologe. 1998;95:176-180.
FULL TEXT
|
ISI
| PUBMED
15. Johnson SM. Neurotrophic corneal defects after diode laser cycloablation. Am J Ophthalmol. 1998;126:725-727.
FULL TEXT
|
ISI
| PUBMED
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
RELATED ARTICLE
A New Role for Diode Laser Transscleral Cyclophotocoagulation?
M. Roy Wilson
Arch Ophthalmol. 2001;119(3):424-425.
EXTRACT
| FULL TEXT
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
A two centre study of the dose-response relation for transscleral diode laser cyclophotocoagulation in refractory glaucoma
Murphy et al.
Br J Ophthalmol 2003;87:1252-1257.
ABSTRACT
| FULL TEXT
Glaucoma in west Africa: a neglected problem
Egbert
Br J Ophthalmol 2002;86:131-132.
ABSTRACT
| FULL TEXT
A New Role for Diode Laser Transscleral Cyclophotocoagulation?
Wilson
Arch Ophthalmol 2001;119:424-425.
FULL TEXT
|