  Vol. 116 No. 9, September 1998 TABLE OF CONTENTS
  Archives
  •  Online Features
  Clinical Sciences
 This Article
 •Abstract
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citation map
 •Citing articles on HighWire
 •Citing articles on ISI (14)
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Glaucoma
 •Ophthalmological Disorders, Other
 •Alert me on articles by topic

Motion Automated Perimetry Identifies Early Glaucomatous Field Defects

Charles F. Bosworth, BA; Pamela A. Sample, PhD; Neeru Gupta, MD, PhD; Renuka Bathija, MD; Robert N. Weinreb, MD

Arch Ophthalmol. 1998;116:1153-1158.

ABSTRACT

Objective  To determine if motion automated perimetry can identify early glaucomatous visual field defects in patients with suspected glaucoma (by disc), those with ocular hypertension, and those with primary open-angle glaucoma.

Methods  Motion automated perimetry, a foveally centered motion test, and standard visual field tests were conducted on one randomly selected eye of normal patients (n=38), patients with suspected glaucoma (by disc) (n=28), patients with ocular hypertension (n=18), and patients with primary open-angle glaucoma (n=21). Subjects' performance on both motion tests were compared with their performance on standard perimetry.

Results  Perimetric motion thresholds significantly distinguished the groups (P<=.001), while the foveally centered motion test was unable to separate them (P<=.32). Of the total patients, 90.5% of those with glaucoma, 39.3% of those with suspected glaucoma, 27.8% of those with ocular hypertension, and 5.3% of the normal subjects had abnormal results on motion automated perimetry testing. Perimetric motion thresholds were significantly correlated with standard visual field thresholds (P<=.001).

Conclusion  Motion automated perimetry identifies visual field defects in patients who already show standard visual field loss as well as in a moderate percentage of those with suspected glaucoma and ocular hypertension, indicating that the testing of discrete locations might be necessary for increased diagnostic utility.



INTRODUCTION
 Jump to Section
 •Top
 •Introduction
 •Subjects and methods
 •Results
 •Comment
 •Author information
 •References

DETECTABLE OPTIC nerve head change or visual field loss may not be observed until there is considerable neural damage; hence, the development of new testing paradigms that can identify patients likely to have or develop glaucoma is of interest.

Psychophysical tests of both the parvocellular and the magnocellular visual systems have indicated abnormalities in patients with glaucoma and suspected glaucoma relative to normal controls.1-3 Magnocellular retinal ganglion cell function has been evaluated with various types of stimuli in patients with glaucoma,4-10 suspected glaucoma, and ocular hypertension (OHT).11-17 The studies evaluating motion perception have indicated that both patients with glaucoma and those with suspected glaucoma with normal standard visual fields often have defects in motion sensitivity. While there is still debate over which type of motion stimulus is the best, testing motion perception with random-dot kinematograms (RDKs) reduces the presence of cues (position and form) that nonmagnocellular ganglion cells can process.18

The purpose of this study was to determine if a perimetric motion test using RDKs in a direction discrimination paradigm, motion automated perimetry (MAP), can differentiate among normal subjects, those with suspected glaucoma (by disc), those with pure OHT, and those with primary open-angle glaucoma, and to compare it with both a separate full-field foveally centered RDK and standard automated perimetry.


SUBJECTS AND METHODS
 Jump to Section
 •Top
 •Introduction
 •Subjects and methods
 •Results
 •Comment
 •Author information
 •References

SUBJECTS

Subjects for this study were patients with primary open-angle glaucoma (n=21), patients with suspected glaucoma (by disc) (n=28), patients with OHT (n=18), and normal controls (n=38). One eye was randomly chosen from each subject. The different subject groups were not significantly different in age (P<=.09). Mean (±SD) ages were 66.3±11.8 years for the glaucoma group, 62.2±11.2 years for the suspected glaucoma group, 57.1±12.7 years for the OHT group, and 62.0±10.3 years for the normal control group.

Patients with primary open-angle glaucoma were defined by the following criteria: (1) open angles with abnormal optic discs based on 1 or more of the following: cup-disc ratio asymmetry between the 2 eyes of 0.2 mm or more, localized or diffuse rim defects, excavation, retinal nerve fiber layer defects, or disc hemorrhages as determined by masked review of simultaneous stereophotographs by a glaucoma expert; (2) previously documented standard visual field loss determined by visual field analysis (program 24-2, Humphrey Visual Field Analyzer 640, Humphrey Instruments, San Leandro, Calif), with corrected pattern SDs outside the 95% confidence limits or glaucoma hemifield test results outside the 99% confidence limits.

Patients with suspected glaucoma were identified by the above criteria for abnormal optic discs. However, they had normal standard visual fields, with corrected pattern SDs within the 95% confidence limits and glaucoma hemifield test results within normal limits.

Normal controls met the following inclusion criteria: (1) normal optic discs based on above criteria; (2) intraocular pressures less than 21 mm Hg on all prior examinations; and (3) normal standard visual fields based on above criteria.

Patients with OHT met all of the above criteria for normal controls except they had intraocular pressures greater than 23 mm Hg on 2 separate examinations.

For this study, visual field reliability indices were set at 25% or less for fixation losses, false-positive errors, and false-negative errors. Mean (±SD) defects on standard perimetry were -7.6±5.61 for eyes with glaucoma, -0.69±1.9 for eyes with suspected glaucoma, 0.26±1.12 for eyes with OHT, and 0.42±1.4 for normal control eyes.

This study was approved by the Human Subjects Committee of the University of California, San Diego, and was undertaken with the informed consent of each subject.

RANDOM-DOT MOTION DISPLAY

The stimulus parameters and design of MAP have been previously reported.4-5 The current stimulus was within known ranges for testing the motion system while still allowing glaucoma patients with impaired vision to perceive the display.

The motion stimulus was produced on a computer-controlled imaging display monitor (Barco, Kennesow, Ga) with 1024x768 lines of resolution and a refresh rate of 75 Hz. Each pixel subtended 0.31 mm (7.35 minutes of arc at the viewing distance of 16.5 cm). The monitor was driven by a Power PC 8100 Macintosh computer (Apple, Cupertino, Calif) using a 24xli video card (RasterOps, Santa Clara, Calif).

Seven frames were shown in rapid succession to create the apparent motion stimulus, which lasted for 420 milliseconds. Within each of these frames, 20 dots were randomly placed within a circular test region of 7.3° of visual angle. These dots moved at a constant velocity of 8.2° per second in random directions. A new direction of motion was chosen after each spatial displacement. Randomly chosen signal dots were displaced together in 1 of 4 cardinal directions (right, left, up, down) to create the coherent motion signal that the patients were to detect. The signal ranged in strength from 0% to 100% coherence. Signal dots remained the same for all 7 frames of the display and had the same spatial displacement as the noise. This aspect of the stimulus makes it differ slightly from a traditional RDK, which randomly selects the signal dots after each displacement. The implications of this signal rule have been discussed in a previous article,5 where it was shown that a signal of this nature embedded in a surround of vectored motion is not significantly influenced by nonmotion cues.

Thresholds were determined by means of a staircase procedure. Staircases began with a coherence value of 80% and a step size of 20% coherence. Each staircase reversal resulted in a halving of the step size down to the minimum of 5% coherence. The staircase was terminated by 3 reversals at the minimum step size. Threshold was taken as the mean of the last 3 reversal points.

We designed the perimetric motion stimulus to test 14 retinal locations because early glaucoma is often characterized by localized peripheral visual field loss. The test locations were arranged in a pattern that mapped the subject's motion thresholds across the central 24° of visual field in approximately 15 minutes (Figure 1). The locations of test were chosen to identify peripheral arcuate and nasal step defects and to overlap with visual field testing locations.



View larger version (31K):
[in this window]
[in a new window]
Figure 1. An example of standard visual field (program 24-2, Humphrey Instruments, San Leandro, Calif) showing the actual threshold values in numeric decibels (left) and the gray scale (right). Motion automated perimetry thresholds are superimposed over the standard gray scale to show the 14 test locations. The number at the origin is threshold for the foveally centered motion test; it was actually much larger than is depicted (diameter, 25.1°). The remaining 14 are in the correct scale (diameter, 7.3°) and show threshold values from the perimetric motion test for this patient.


The stimulus for the foveally centered motion test was centered on the fovea and subtended a circular region 25.1° in diameter with a dot density identical to that used in the smaller perimetric displays. All other parameters remained the same.

TEST PROCEDURES

The patient sat in a darkened room with a patch placed over the nontest eye. He or she rested in a chin rest while viewing the screen through proper refraction for the test distance of 16.5 cm.

The patient focused on a black fixation "X" in the center of the display and adapted for 2 minutes to the background illumination. Fixation was monitored by the test administrator on a separate video display system. Trials where fixation was lost were aborted and retested later in the program. The testing procedure was then explained to the subject.

The session began with a foveal practice test, which was identical to the foveally centered single-stimulus test. The patient's performance on these trials was observed by the test administrator to make sure that the task was properly understood. After completing the foveal practice, the patient either received the larger foveally centered test or the 14-location perimetric motion test. The order of presentation for the 2 types of motion stimuli was randomized.

For all testing, the patient verbally reported the perceived direction (left, right, up, or down) in which the dots were moving in a 4-alternative forced-choice paradigm. The test administrator pressed an arrow key corresponding to the direction indicated by the subject. The patient could respond at any time during or after the stimulus presentation. The program gave a 2-second delay before presenting the next stimulus. Subjects were informed that they could pause and rest at any time during the test. The whole procedure lasted approximately 30 minutes.

STATISTICAL ANALYSES

The perimetric threshold values averaged across the 14 test points (whole field), the perimetric threshold values averaged across the 7 superior field test points (superior field), the perimetric threshold values averaged across the 7 inferior field test points (inferior field), and motion coherence thresholds for the foveally centered motion test were the main dependent measures of this study. Analysis of variance was performed on all dependent measures to indicate if a significant difference existed in motion thresholds among the normal, suspected glaucoma, OHT, and glaucoma subjects. If a significant difference was indicated, a Tukey-Kramer honestly significant difference test was performed to localize the effect.

The Pearson r was computed for the correlation of the mean defects on MAP and standard perimetry. Mean defect on standard perimetry was computed by averaging the patient's threshold values for standard visual field test points that overlapped with each of the MAP test locations to create 14 values. These values were subtracted from the normal averages for each test point separately. The obtained difference scores for all 14 points were averaged to give an overall mean defect on standard perimetry for each patient. Mean defect on MAP was the average of 14 difference scores from normal.

An abnormal MAP field was defined as having 3 or more motion threshold values at least 2 SDs above normal with at least 2 of the abnormal points adjacent to each other. This is a conservative definition because 3 motion locations would cover at least 8 standard visual field locations.

Receiver operator characteristic (ROC) analysis was performed on the percent coherence variables for the glaucoma patients. Receiver operator characteristic curves were plotted for sensitivity vs 1 - specificity for possible cutoff points from 0% to 100% coherence. Sensitivity is the ability to correctly identify glaucoma. Specificity is the ability to correctly confirm a normal field. To quantify the ability of each measure to separate the normal subjects and those with glaucoma, we calculated the percentage of area under the ROC curve. This was accomplished by fitting a function to the ROC from the point (x=100, y=100) in the upper left corner to the first point where x=0 (Figure 2). A fit to the data points was accepted if more than 95% of the variance between the points could be accounted for by the function. The area under the curve was determined by integrating the function from x=0 to x=100 with the x -axis defining the lower bound of the integration area. Because the total area possible is known, the percent area is the area under the curve divided by 100. Using this measure, an area under the ROC curve of 50% represents chance while an area under the ROC curve of 100% represents perfect discrimination.



View larger version (20K):
[in this window]
[in a new window]
Figure 2. Receiver operator characteristic analysis results for the glaucoma patients' mean motion coherence thresholds for the foveally centered motion test, motion automated perimetry averaged across the 14 test points (whole field), motion automated perimetry averaged across the 7 superior field test points (superior field), and motion automated perimetry averaged across the 7 inferior field test points (inferior field). The line from (0, 0) to (100, 100) represents chance performance. The curves are the best-fit functions for each average accounting for more than 95% of the variance. Each function was integrated from x=0 to x =100 to give an estimate of the area under the curve for the 3 measures separately.


Optimal general error rates were computed for the different measures. The general error rate is the percentage of subjects misclassified when a specific cutoff value in percent coherence is selected. Therefore, choosing a cutoff that minimized the general error rate represents a balance between sensitivity and specificity such that the lowest percentage of misses and false alarms possible occurs. The sensitivity, specificity, and optimal cutoff for the different measures are reported at the lowest general error rate.


RESULTS
 Jump to Section
 •Top
 •Introduction
 •Subjects and methods
 •Results
 •Comment
 •Author information
 •References

Analysis of variance indicated that the foveally centered motion test was unable to differentiate the patients with glaucoma, suspected glaucoma, or OHT from normal controls (P<=.32) while MAP was able to do so (Table 1). Additional pairwise comparisons on the perimetric results indicated that the patients with glaucoma and those with suspected glaucoma were significantly different from normal controls throughout the visual field (Table 1). Considering the means for the foveally centered motion test, however, a nonsignificant trend is apparent with increasing coherence thresholds for the OHT and suspected glaucoma groups compared with normal controls. Using an {alpha} level of .05, the power of the foveally centered test was relatively low (0.32), with 233 observations required to obtain a significant result. This is in contrast to the 14 subjects required for a significant result when the superior field average from the perimetric motion test was used in the power calculations. Because the mean defect for patients with glaucoma on standard perimetry ranged from -0.57 to -20.26, the analysis was redone excluding all mean defects greater than -6.0 dB to give an evaluation of MAP for detection of early glaucoma (n=9). The patients with early glaucoma remained significantly different from normal controls (Table 1).


View this table:
[in this window]
[in a new window]
Table 1. Mean (±SD) Motion Coherence Thresholds for All Patients and Patients With a Mean Defect Less Than -6.0 dB


Although MAP did not differentiate the patients with OHT from normal controls, it did identify 5 (27.8%) of the 18 patients with OHT, 19 (90.5%) of the 21 patients with glaucoma, 11 (39.3%) of the 28 patients with suspected glaucoma, and 2 (5.3%) of the 38 normal subjects as abnormal on the examination.

Receiver operator characteristic analysis indicated that the superior field average from MAP gave the largest area under the ROC curve (98.2%). The general error rate for the superior field average was 5.7%, yielding a specificity of 97.4% and a sensitivity of 91.3%. The larger foveally centered stimulus scored the poorest, with 62.4% of the total area under the ROC curve. See Table 2 for a full listing of the ROC results.


View this table:
[in this window]
[in a new window]
Table 2. Receiver Operator Characteristic Analysis Results for the Foveal Motion Test and Motion Automated Perimetry Comparing Eyes With Glaucoma With Normal Eyes


A significant correlation between all patients' standard visual field and MAP mean defects was found (r2=0.35, P<.001) (Figure 3).



View larger version (13K):
[in this window]
[in a new window]
Figure 3. This graph shows the correlation between the mean defect on standard automated perimetry and the mean defect on motion automated perimetry across the patients with glaucoma, ocular hypertension, and suspected glaucoma (r2=0.35, P <.001).


Individual patient data are shown as illustrations. Glaucoma patients 1 (Figure 4, top) and 3 (Figure 4, bottom) show a good correspondence between standard visual fields and MAP. Patient 3 was one of the 2 glaucoma patients who were not identified by MAP because 3 test points were not 2 SDs above normal. Patient 2 (Figure 4, center) shows a more extensive defect on MAP than is present on standard automated perimetry.



View larger version (37K):
[in this window]
[in a new window]
Figure 4. Pattern deviation plots for standard automated perimetry (left side) showing the match with their results on motion automated perimetry (right side) in 3 patients with glaucoma. For motion automated perimetry, the number at the origin is threshold for the foveally centered motion test, while the remaining 14 numbers are threshold values from motion automated perimetry.


Figure 5 shows the results for 3 patients at risk for developing primary open-angle glaucoma. Patients 5 and 7 had suspected glaucoma. Patient 6 had OHT. All 3 patients' standard visual fields were normal while their MAP results showed clear defects.



View larger version (39K):
[in this window]
[in a new window]
Figure 5. Pattern deviation plots for standard automated perimetry (left side) showing the match with their results on motion automated perimetry (right side) in 2 patients with suspected glaucoma5, 7 and 1 patient with ocular hypertension.6 For motion automated perimetry, the number at the origin is threshold for the foveally centered motion test, while the remaining 14 numbers are threshold values from motion automated perimetry.


It should be noted that in Figure 4 and Figure 5 the pattern deviation plot (STATPAC II analysis, Humphrey Instruments) was used for comparison to motion threshold values. We chose to use pattern deviation because it reduces the effects on the visual field due to general reduction in sensitivity from optical factors, such as blurring, cataract, and pupil size. Because MAP is resilient to the confounding influences of blurring, cataract, and pupil size,14, 19 using the pattern deviation makes it more likely that the results are due to glaucomatous damage and thereby more comparable to MAP.


COMMENT
 Jump to Section
 •Top
 •Introduction
 •Subjects and methods
 •Results
 •Comment
 •Author information
 •References

Early attempts to measure motion discrimination thresholds with RDKs used large centrally presented fields of dots and produced mixed results. Silverman et al12 found that motion thresholds were elevated in patients with glaucoma and OHT. This finding was later replicated by Trick et al,14 but could not be replicated by Bullimore et al7 and Graham et al.10 The lack of sensitivity to glaucoma reported in these studies could have resulted from either their long presentation durations (>1 second) or their large stimulus aperture sizes (>19°). Long display durations could provide even a damaged motion system with enough time to integrate a noisier or weaker motion signal. Large aperture sizes could allow both spared and nonspared regions of the visual field to contribute to a motion percept.

Consistent with the theory that larger aperture sizes might have lower diagnostic power, perimetric motion tests using small localized targets have had considerable success in detecting glaucomatous visual field loss. Using RDKs with variable aperture sizes in a detection paradigm, Wall and colleagues have reported significant elevations in perimetric size thresholds for patients with glaucoma9 and OHT.17 Similar results using RDKs in perimetric direction discrimination tasks rather than detection tasks have been reported in patients with glaucoma3-6 and those with suspected glaucoma.5 A pure OHT group has not previously been assessed using RDKs in a perimetric direction discrimination paradigm. This could have clinical significance as the sensitivity of direction discrimination tasks to glaucomatous visual field loss might exceed that of motion detection tasks.8

In a previous study to determine if glaucoma-related loss of motion sensitivity is localized in retinal locus, we presented a discrete motion target at a location of standard visual field loss and a location of standard visual field sparing in the same eye of a glaucoma patient.4 The 2 test locations mirrored each other across the horizontal midline. The results indicated a significant difference between spared and nonspared retinal locations, suggesting that glaucoma causes localized loss of motion sensitivity.4

The ROC results of the present study further confirm that a foveally presented RDK is less sensitive to glaucomatous visual field loss than MAP, which tests discrete retinal locations. This becomes apparent when comparing the percentage of area under the ROC curves. The greatest area under the ROC curve and therefore the greatest sensitivity was found for the superior half of the MAP field.

Motion automated perimetry can also identify field loss in a percentage of the patients with suspected glaucoma and those with OHT, suggesting that loss of motion sensitivity may precede standard visual field loss. The percentage of patients with OHT identified as abnormal is consistent with previous work using RDKs in a detection paradigm17 and indicates that MAP visual field abnormality might exist prior to detectable abnormality of the optic nerve.

While the stimulus design used in this study was successful in separating glaucomatous from normal eyes, more work is necessary to evaluate possible limitations of the current test design. A stimulus duration of 420 milliseconds necessitates careful fixation monitoring in the administration of MAP. A second possible limitation is normal variability. The normal variability creates a relatively high SD (between 8.0% and 11.6% coherence) (Table 1). Future work using direction discrimination tasks should assess test-retest variability and ways to reduce it.

Longitudinal follow-up with MAP and conventional automated perimetry is required to determine if the eyes with OHT and suspected glaucoma with MAP defects will eventually develop standard visual field loss.


AUTHOR INFORMATION
 Jump to Section
 •Top
 •Introduction
 •Subjects and methods
 •Results
 •Comment
 •Author information
 •References

Accepted for publication June 3, 1998.

This study was supported by grant EY-08208 from the National Eye Institute, Bethesda, Md (Dr Sample), and by the Samuel E. McLaughlin Foundation of Canada, Toronto, Ontario (Dr Gupta).

Corresponding author: Pamela A. Sample, PhD, Department of Ophthalmology-0946, University of California, San Diego, La Jolla, CA 92093-0946 (e-mail: psample{at}eyecenter.ucsd.edu).

From the Glaucoma Center and Department of Ophthalmology, University of California, San Diego (Mr Bosworth and Drs Sample, Gupta, Bathija, and Weinreb); and Department of Ophthalmology, University of Toronto, Toronto, Ontario (Dr Gupta).


REFERENCES
 Jump to Section
 •Top
 •Introduction
 •Subjects and methods
 •Results
 •Comment
 •Author information
 •References

1. Sample PA, Madrid ME, Weinreb RN. Evidence for a variety of functional defects in glaucoma-suspect eyes. J Glaucoma. 1994;3(suppl):S5-S18.
2. Casson EJ, Johnson CA, Shapiro LR. Longitudinal comparison of temporal-modulation perimetry with white-on-white and blue-on-yellow perimetry in ocular hypertension and early glaucoma. J Opt Soc Am. 1993;10:1792-1806.
3. Sample PA, Bosworth CF, Weinreb RN. Short-wavelength automated perimetry and motion automated perimetry in glaucoma. Arch Ophthalmol. 1997;115:1129-1133. ABSTRACT
4. Bosworth CF, Sample PA, Weinreb RN. Motion perception thresholds in areas of glaucomatous visual field loss. Vision Res. 1997;37:355-364. FULL TEXT | ISI | PUBMED
5. Bosworth CF, Sample PA, Weinreb RN. Perimetric motion thresholds are elevated in glaucoma suspects and glaucoma patients. Vision Res. 1997;37:1989-1997. FULL TEXT | ISI | PUBMED
6. Joffe KM, Raymond JE, Crichton A. Motion coherence perimetry in glaucoma and suspected glaucoma. Vision Res. 1997;37:955-964. FULL TEXT | ISI | PUBMED
7. Bullimore MA, Wood JM, Swenson K. Motion perception in glaucoma. Invest Ophthalmol Vis Sci. 1993;34:3526-3533. FREE FULL TEXT
8. Sahraie A, Burber JL, Edgar DF, Weiskrantz L. Motion discrimination of single targets: comparison of preliminary findings in normal subjects and patients with glaucoma. Graefs Arch Clin Exp Ophthalmol. 1996;234:553-560. FULL TEXT | ISI | PUBMED
9. Wall M, Ketoff KM. Random dot motion perimetry in patients with glaucoma and in normal subjects. Am J Ophthalmol. 1995;120:587-596. ISI | PUBMED
10. Graham SL, Drance SM, Chauhan BC, et al. Comparison of psychophysical and electrophysiological testing in early glaucoma. Invest Ophthalmol Vis Sci. 1996;37:2651-2662. FREE FULL TEXT
11. Fitzke F, Poinoosawmy D, Nagasubramanian S, Hitchings RA. Peripheral displacement thresholds in glaucoma and ocular hypertension. In: Heijl A, ed. Perimetry Update. Amsterdam, the Netherlands: Kugler & Ghedini Publications; 1989:399-405.
12. Silverman SE, Trick GL, Hart WM Jr. Motion perception is abnormal in primary open-angle glaucoma and ocular hypertension. Invest Ophthalmol Vis Sci. 1990;31:722-729. FREE FULL TEXT
13. Watkins R, Buckingham T. Motion perception hyperacuity is abnormal in primary open-angle glaucoma and ocular hypertension [abstract]. Invest Ophthalmol Vis Sci. 1991;32(suppl):1103.
14. Trick GL, Steinman SB, Amyot M. Motion perception deficits in glaucomatous optic neuropathy. Vision Res. 1995;35:2225-2233. FULL TEXT | ISI | PUBMED
15. Johnson CA, Marshall D, Eng KM. Displacement threshold perimetry in glaucoma using a Macintosh computer system and a 21-inch monitor. In: Mills RP, Wall M, eds. Perimetry Update. Amsterdam, the Netherlands: Kugler Publications; 1995:103-110.
16. Scholl HPN, Zrenner E. Motion thresholds of colored stimuli of different luminance contrasts are increased in ocular hypertension and early primary-open-angle-glaucoma (POAG). Color Vision Deficiencies. 1995;13:73-85.
17. Wall M, Jennisch CS, Munden PM. Motion perimetry identifies nerve fiber bundlelike defects in ocular hypertension. Arch Ophthalmol. 1997;115:26-33. ABSTRACT
18. Nakayama K, Tyler CW. Psychophysical isolation of movement sensitivity by removal of familiar position cues. Vision Res. 1981;21:427-433. FULL TEXT | ISI | PUBMED
19. Turano K, Wang X. Motion thresholds in retinitis pigmentosa. Invest Ophthalmol Vis Sci. 1992;33:2411-2422. FREE FULL TEXT


THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES

Sources of Motion-Sensitivity Loss in Glaucoma
Falkenberg and Bex
IOVS 2007;48:2913-2921.
ABSTRACT | FULL TEXT  

The Relationship between Nerve Fiber Layer and Perimetry Measurements
Harwerth et al.
IOVS 2007;48:763-773.
ABSTRACT | FULL TEXT  

Probing glaucoma visual damage by rarebit perimetry
Brusini et al.
Br. J. Ophthalmol. 2005;89:180-184.
ABSTRACT | FULL TEXT  

Detecting Early Glaucoma by Assessment of Retinal Nerve Fiber Layer Thickness and Visual Function
Bowd et al.
IOVS 2001;42:1993-2003.
ABSTRACT | FULL TEXT  

Visual Function-Specific Perimetry for Indirect Comparison of Different Ganglion Cell Populations in Glaucoma
Sample et al.
IOVS 2000;41:1783-1790.
ABSTRACT | FULL TEXT  

Loss of Neurons in Magnocellular and Parvocellular Layers of the Lateral Geniculate Nucleus in Glaucoma
Yucel et al.
Arch Ophthalmol 2000;118:378-384.
ABSTRACT | FULL TEXT  