Scleral Changes During Diathermy ApplicationInfluence of Electrode Type
Henri F. deGuillebon, Dipl Eng;
Yasuo Ishii
Arch Ophthalmol. 1970;83(6):752-759.
Abstract
Three degrees of scleral damage were observed when heating sclera in physiological saline: formation of vacuoles at 50 C, fragmentation of collagen bundles at 59 C, and denaturation of collagen at 65 C. The same degrees of damage were found after diathermy applications. The gradient of temperatures indicated by the silver-stained cryosections demonstrates the pathway of the electric current through sclera. A point electrode produces high temperatures at the point of contact with sclera. With a flat electrode, the electric current flows along the periphery of the metallic electrode and scleral damage is more prominent along the edges of the electrode. Scleral damage can be decreased with a cuff of saline solution around the electrode.
Author Affiliations
Boston
From the Department of Retina Research, Retina Foundation, Boston.
Footnotes
Submitted for publication Oct 7, 1969.
Reprint requests to Library, Retina Foundation, 100 Charles River Plaza, Boston 02114 (Mr. deGuillebon).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Experimental Transconjunctival Diode Laser Retinal Photocoagulation Through Silicone Scleral Exoplants
Nanda and Han
Arch Ophthalmol 1995;113:926-931.
ABSTRACT
Microwave Diathermy of the Retina and Choroid
Finger et al.
Arch Ophthalmol 1993;111:254-258.
ABSTRACT
An Insulated Diathermy Electrode
El-Mofty and Schepens
Arch Ophthalmol 1980;98:1115-1116.
ABSTRACT
Scleral Bursting Strength: Results After Treatment With Ruby Laser, Cryotherapy, or Diathermy
Beckman et al.
Arch Ophthalmol 1975;93:428-428.
ABSTRACT
|