Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle Glaucoma
Matthias C. Grieshaber, MD;
Andreas Schoetzau, MS;
Claudia Zawinka, MD;
Josef Flammer, MD;
Selim Orgul, MD
Arch Ophthalmol. 2007;125(6):740-744.
Objective To compare the dependence of dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) on central corneal thickness (CCT) in primary open-angle glaucoma.
Methods In a prospective study, the interocular (right vs left eye) difference in intraocular pressure measured by DCT and GAT was compared with the interocular CCT difference in 125 patients with primary open-angle glaucoma.
Results Dynamic contour tonometry measurements (mean ± SD, 19.4 ± 4.1 mm Hg) were significantly (P = .004) higher than GAT measurements (mean ± SD, 15.5 ± 3.4 mm Hg), correlating significantly with each other (r2 = 0.82, P<.001). The interocular difference in intraocular pressure correlated significantly with the interocular CCT difference for GAT (r = 0.30, P = .001) and DCT (r = 0.23, P = .02) readings. Dynamic contour tonometry and GAT intraocular pressure differences significantly increased with older age (slope, 0.033 [95% confidence interval, 0.002-0.064] mm Hg/y; P =.03) but not with thicker CCT (slope, 0.006 [95% confidence interval, –0.003 to 0.017] mm Hg/µm; P =.22).
Conclusions In this series, GAT and DCT measurements were dependent on CCT in patients with primary open-angle glaucoma. Because intraocular pressure differences between DCT and GAT were independent of CCT, DCT and GAT are susceptible to similar measurement biases depending on CCT.
Author Affiliations: Department of Ophthalmology, University Hospital Basel, Basel, Switzerland.
|