You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 125 No. 6, June 2007 TABLE OF CONTENTS
  Archives
  •  Online Features
  Clinical Sciences
 This Article
 •Full text
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citation map
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Corneal Disorders
 •Glaucoma
 •Alert me on articles by topic

Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle Glaucoma

Matthias C. Grieshaber, MD; Andreas Schoetzau, MS; Claudia Zawinka, MD; Josef Flammer, MD; Selim Orgul, MD

Arch Ophthalmol. 2007;125(6):740-744.

Objective  To compare the dependence of dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) on central corneal thickness (CCT) in primary open-angle glaucoma.

Methods  In a prospective study, the interocular (right vs left eye) difference in intraocular pressure measured by DCT and GAT was compared with the interocular CCT difference in 125 patients with primary open-angle glaucoma.

Results  Dynamic contour tonometry measurements (mean ± SD, 19.4 ± 4.1 mm Hg) were significantly (P = .004) higher than GAT measurements (mean ± SD, 15.5 ± 3.4 mm Hg), correlating significantly with each other (r2 = 0.82, P<.001). The interocular difference in intraocular pressure correlated significantly with the interocular CCT difference for GAT (r = 0.30, P = .001) and DCT (r = 0.23, P = .02) readings. Dynamic contour tonometry and GAT intraocular pressure differences significantly increased with older age (slope, 0.033 [95% confidence interval, 0.002-0.064] mm Hg/y; P =.03) but not with thicker CCT (slope, 0.006 [95% confidence interval, –0.003 to 0.017] mm Hg/µm; P =.22).

Conclusions  In this series, GAT and DCT measurements were dependent on CCT in patients with primary open-angle glaucoma. Because intraocular pressure differences between DCT and GAT were independent of CCT, DCT and GAT are susceptible to similar measurement biases depending on CCT.


Author Affiliations: Department of Ophthalmology, University Hospital Basel, Basel, Switzerland.