Agreement Between Gonioscopy and Ultrasound Biomicroscopy in Detecting Iridotrabecular Apposition
Yaniv Barkana, MD;
Syril K. Dorairaj, MD;
Yariv Gerber, PhD;
Jeffrey M. Liebmann, MD;
Robert Ritch, MD
Arch Ophthalmol. 2007;125(10):1331-1335.
Objective To assess the agreement between findings obtained at dark-room gonioscopy and ultrasound biomicroscopy (UBM) in the diagnosis of iridotrabecular apposition in light and dark conditions.
Methods We enrolled patients with appositional angle closure at dark-room gonioscopy performed using a 1-mm slitlamp beam that did not cross the pupil. Ultrasound biomicroscopic images were acquired in normal room light and subsequently with all room lights off. Images were evaluated for the presence or absence of iris-cornea contact. The angle opening distance at 500 µm was calculated.
Results Iridotrabecular apposition in at least 1 angle quadrant was demonstrated in all 18 eyes at dark-room gonioscopy, 17 eyes (94%) at dark-room UBM, and only 10 eyes (56%) at UBM in room light. Of 18 superior angles that were appositionally closed at dark-room gonioscopy, apposition was demonstrated on UBM images in 16 (89%) in a dark room but only 6 (33%) in room light. Angle opening distance was less during dark-room gonioscopy in all but the nasal quadrant.
Conclusion We found high agreement between gonioscopy and UBM when both are performed in a completely dark room. Our findings support the recommendation that, in routine clinical practice, gonioscopy be performed in a dark room to avoid misdiagnosis of treatable iridotrabecular apposition.
Author Affiliations: Glaucoma Unit, Department of Ophthalmology, Assaf Harofeh Medical Center, Zerifin, Israel (Dr Barkana); Glaucoma Service, Einhorn Clinical Research Center, New York Eye and Ear Infirmary, New York (Drs Dorairaj and Ritch); Department of Epidemiology and Preventive Medicine, Sackler Medical School, Tel Aviv University, Tel Aviv, Israel (Dr Gerber); Department of Ophthalmology, Manhattan Eye, Ear and Throat Hospital, New York, New York (Dr Liebmann); Department of Ophthalmology, New York University Medical Center, New York (Dr Liebmann); and Department of Ophthalmology, New York Medical College, Valhalla (Dr Liebmann).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
RELATED ARTICLE
Shedding Light on Gonioscopy
Paul Palmberg
Arch Ophthalmol. 2007;125(10):1417-1418.
EXTRACT
| FULL TEXT
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Anatomical Changes of the Anterior Chamber Angle With Anterior-Segment Optical Coherence Tomography
Liu
Arch Ophthalmol 2008;126:1682-1686.
ABSTRACT
| FULL TEXT
Shedding Light on Gonioscopy
Palmberg
Arch Ophthalmol 2007;125:1417-1418.
FULL TEXT
|