Accuracy and Reliability of Remote Retinopathy of Prematurity Diagnosis
Michael F. Chiang, MD, MA;
Jeremy D. Keenan, MD, MPH;
Justin Starren, MD, PhD;
Yunling E. Du, PhD;
William M. Schiff, MD;
Gaetano R. Barile, MD;
Joan Li, MD;
Rose Anne Johnson, RN;
Ditte J. Hess, CRA;
John T. Flynn, MD
Arch Ophthalmol. 2006;124:322-327.
Objective To determine the accuracy and reliability of retinopathy of prematurity (ROP) diagnosis using remote review of digital images by 3 masked ophthalmologist readers.
Methods An atlas was compiled of 410 retinal photographs from 163 eyes of 64 low-birth-weight infants taken using a wide-angle digital fundus camera. All the images were independently reviewed by 3 readers, and the diagnosis in each eye was classified into 1 of 4 ordinal categories: no ROP, mild ROP, type 2 prethreshold ROP, or ROP requiring treatment. Findings were compared with a reference standard of dilated indirect ophthalmoscopy with scleral depression performed by an experienced pediatric ophthalmologist.
Results Sensitivities/specificities of the diagnosis of any ROP were 0.845/0.910 for the first reader, 0.816/0.955 for the second reader, and 0.864/0.493 for the third reader. Sensitivities/specificities of the diagnosis of ROP requiring treatment were 0.850/0.960 for the first reader, 0.850/0.973 for the second reader, and 0.900/0.953 for the third reader. When ROP was classified into ordinal categories, the overall weighted for interreader reliability was 0.743. Intrareader reliability for detection of low-risk prethreshold ROP or worse was 100% for all readers.
Conclusion The accuracy, interreader reliability, and intrareader reliability of remote diagnosis of clinically relevant ROP based on digital imaging are substantial.
Author Affiliations: Departments of Ophthalmology (Drs Chiang, Keenan, Schiff, Barile, Li, and Flynn), Biomedical Informatics (Drs Chiang and Starren), and Radiology (Dr Starren), Columbia University College of Physicians and Surgeons, New York, NY; Department of Epidemiology and Population Health, Albert Einstein College of Medicine, New York (Dr Du); Department of Pediatric Nursing, Jackson Memorial Hospital, Miami, Fla (Ms Johnson); and Bascom Palmer Eye Institute, Miami (Ms Hess).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
 |
Technological Advances in Pediatric Eye Care
Wallace et al.
Arch Ophthalmol 2009;127:805-806.
FULL TEXT
Stanford University Network for Diagnosis of Retinopathy of Prematurity (SUNDROP): 12-month experience with telemedicine screening
Murakami et al.
Br J Ophthalmol 2008;92:1456-1460.
ABSTRACT
| FULL TEXT
Systematic Review of Digital Imaging Screening Strategies for Retinopathy of Prematurity
Kemper et al.
Pediatrics 2008;122:825-830.
ABSTRACT
| FULL TEXT
Variation in Appearance of Severe Zone 1 Retinopathy of Prematurity During Wide-angle Contact Photography
Koreen et al.
Arch Ophthalmol 2008;126:736-737.
FULL TEXT
Cost-Utility Analysis of Telemedicine and Ophthalmoscopy for Retinopathy of Prematurity Management
Jackson et al.
Arch Ophthalmol 2008;126:493-499.
ABSTRACT
| FULL TEXT
Progression to Severe Retinopathy Predicted by Retinal Vessel Diameter Between 31 and 34 Weeks of Postconception Age
Rabinowitz et al.
Arch Ophthalmol 2007;125:1495-1500.
ABSTRACT
| FULL TEXT
Accuracy of ROPtool vs Individual Examiners in Assessing Retinal Vascular Tortuosity
Wallace et al.
Arch Ophthalmol 2007;125:1523-1530.
ABSTRACT
| FULL TEXT
Telemedical Retinopathy of Prematurity Diagnosis: Accuracy, Reliability, and Image Quality
Chiang et al.
Arch Ophthalmol 2007;125:1531-1538.
ABSTRACT
| FULL TEXT
Interexpert Agreement of Plus Disease Diagnosis in Retinopathy of Prematurity
Chiang et al.
Arch Ophthalmol 2007;125:875-880.
ABSTRACT
| FULL TEXT
Remote image based retinopathy of prematurity diagnosis: a receiver operating characteristic analysis of accuracy
Chiang et al.
Br J Ophthalmol 2006;90:1292-1296.
ABSTRACT
| FULL TEXT
|