Clinical Comparison of Contour and Applanation Tonometry and Their Relationship to Pachymetry
Christoph Kniestedt, MD;
Shan Lin, MD;
Joyce Choe, MD;
Alan Bostrom, PhD;
Michelle Nee, MD;
Robert L. Stamper, MD
Arch Ophthalmol. 2005;123:1532-1537.
Objectives To compare intraocular pressure readings of recently introduced dynamic contour tonometry (DCT) with pneumatonometry (PTG) and Goldmann applanation tonometry (GAT) and to correlate central corneal thickness (CCT) with these readings.
Design Prospective, cross-sectional observation and instrument validation study. We included 258 independent eyes with normal anterior segment examinations results, irrespective of glaucoma diagnosis or glaucoma suspect. After pachymetry, DCT, PTG, and GAT were performed in a randomized order. Intraocular pressures as measured by DCT, PTG, and GAT were compared with each other and with CCT.
Results Eyes with thinner CCTs tended to yield lower intraocular pressure measurements by GAT. A significant correlation (Pearson product moment correlation, P<.001) between CCT and GAT was found with a regression of 0.25 mm Hg per 10 µm (R2 = 0.060). Variation of CCT had no significant effect on intraocular pressure measurements by PTG (P = .10; R2 = 0.01) and DCT (P = .80; R2<0.01). A piecewise regression model showed that GAT readings are not linearly correlated with CCT. Comparison of the slopes below and above 535 µm showed the highest significance (P<.001).
Conclusions Goldmann applanation tonometry readings are potentially influenced by CCT, whereas PTG and DCT seem to be less dependent on CCT. Correlation between CCT and GAT is not linear. A simple correction formula suggesting a linear relationship might not be correct.
Author Affiliations: Departments of Ophthalmology (Drs Kniestedt, Lin, Choe, Nee, and Stamper) and Epidemiology and Biostatistics (Dr Bostrom), University of CaliforniaSan Francisco; and Departments of Ophthalmology, Cantonal Hospital Winterthur, Winterthur, Switzerland (Dr Kniestedt), and University Hospital Zurich, Zurich, Switzerland (Dr Kniestedt).
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Dynamic Contour Tonometry in Comparison to Intracameral IOP Measurements
Boehm et al.
IOVS 2008;49:2472-2477.
ABSTRACT
| FULL TEXT
Effect of Central Corneal Thickness on Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Primary Open-angle Glaucoma
Grieshaber et al.
Arch Ophthalmol 2007;125:740-744.
ABSTRACT
| FULL TEXT
Ocular Pulse Amplitude in Healthy Subjects as Measured by Dynamic Contour Tonometry.
Kaufmann et al.
Arch Ophthalmol 2006;124:1104-1108.
ABSTRACT
| FULL TEXT
|