Precision and Safety of the Pulsed Electron Avalanche Knife in Vitreoretinal Surgery
Jason M. Miller, MSc;
Daniel V. Palanker, PhD;
Alexander Vankov, PhD;
Michael F. Marmor, MD;
Mark S. Blumenkranz, MD
Arch Ophthalmol. 2003;121:871-877.
Background We have developed a new surgical instrument, called the pulsed electron avalanche knife (PEAK; Carl Zeiss Meditec, Jena, Germany), for precise, "cold," and tractionless dissection of tissue in liquid media.
Objective To evaluate the 3-dimensional damage zone induced by the PEAK compared with 2 other standard intraocular surgical instruments, diathermy and retinal scissors.
Methods Damage zone and minimum safe distance were measured in vitro on chick chorioallantoic membrane and in vivo on rabbit retina with the use of propidium iodide staining.
Results The PEAK produced a paracentral zone of cellular structure disruption surrounding a crater and a peripheral zone of structurally intact but abnormally permeable cells. The instrument induced a damage radius that varied from 55 to 300 µm for the range of voltages and pulses typically used during surgery. For comparison, damage radius for microsurgical scissors was 50 µm, and for diathermy, 400 to 850 µm. The PEAK also damaged tissue up to 1.4 mm away by the creation of water flow that formed at the tip of convex probes during collapse of a cavitation bubble. Concave probes, which prevent formation of the water jet, eliminated this effect.
Conclusions The PEAK operated well within accept-able safety limits and may greatly facilitate both posterior segment surgeries (eg, membrane dissection and sheathotomy) and anterior segment procedures (eg, capsulotomy, nonpenetrating trabeculectomy, and iridectomy).
From the Department of Ophthalmology, School of Medicine (Mr Miller and Drs Palanker, Marmor, and Blumenkranz) and W. W. Hansen Experimental Physics Laboratory (Drs Palanker and Vankov), Stanford University, Stanford, Calif. Stanford University is the owner of a patent covering the use of the pulsed electron avalanche knife. Should Stanford University receive royalties or other financial remuneration in the future related to the patent, Dr Palanker may receive a share in accordance with the Stanford University Institutional Patent Policy and Procedures, which include royalty-sharing provisions. Drs Palanker and Blumenkranz are consultants to Carl Zeiss Meditec, Jena, Germany.
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Pulsed electron avalanche knife: new technology for cataract surgery
Priglinger et al.
Br. J. Ophthalmol. 2007;91:949-954.
ABSTRACT
| FULL TEXT
Gene transfer to rabbit retina with electron avalanche transfection.
Chalberg et al.
IOVS 2006;47:4083-4090.
ABSTRACT
| FULL TEXT
Pulsed Electron Avalanche Knife (PEAK-fc) for Dissection of Retinal Tissue
Priglinger et al.
Arch Ophthalmol 2005;123:1412-1418.
ABSTRACT
| FULL TEXT
|