Clinical Investigation of a True Color Scanning Laser Ophthalmoscope
Ayyakkannu Manivannan, PhD;
Jan Van der Hoek, FRCOphth;
Pedro Vieira, PhD;
Allison Farrow, ABIPP;
John Olson, MRCP(UK);
Peter F. Sharp, PhD;
John V. Forrester, MD
Arch Ophthalmol. 2001;119:819-824.
Objective To show true color scanning laser ophthalmoscope (SLO) images produced by simultaneously imaging the retina with red, green, and blue lasers.
Methods Low-power red, green, and blue lasers were combined using fiber optics. By rapidly pulsing the lasers, each point on the fundus is illuminated by the 3 colors in quick succession, with the total power level being similar to that from a single laser. The reflected light is then decoded to extract the red, green, and blue color information and the true color fundus image is displayed live on a computer monitor.
Results Comparison was made between the color SLO images from 5 patients and their digitized fundus photographs. The background fundus and retinal vasculature showed a similar appearance. The SLO gave better quality information in patients with ocular histoplasmosis, macular dystrophy, and optic disc drusen. By operating the color SLO in the indirect mode, macular edema could be clearly seen as lines and ridges surrounding the fovea.
Conclusion The color SLO offers all the advantages of the present commercially available monochromatic device, with the added advantage of true color representation of the fundus without increasing either imaging time or the level of exposure.
From the Departments of Biomedical Physics and Bioengineering (Drs Manivannan, Vieira, and Sharp), Ophthalmology (Drs Van der Hoek, Olson, and Forrester), and Medical Illustration (Ms Farrow), University of Aberdeen, Foresterhill, Aberdeen, Scotland.
RELATED ARTICLE
Archives of Ophthalmology Reader's Choice: Continuing Medical Education
Arch Ophthalmol. 2001;119(6):931-932.
FULL TEXT
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Omitted References on Color Scanning Laser Ophthalmoscopy
Ashman et al.
Arch Ophthalmol 2002;120:1601-1601.
FULL TEXT
A False Use of "True Color"
Bartsch et al.
Arch Ophthalmol 2002;120:675-676.
FULL TEXT
|