An Epidemic of Corneal Destruction Caused by Plasma Gas Sterilization
Rosemary E. Duffy, DDS, MPH;
Sally E. Brown, BSN, MPH;
Kathleen L. Caldwell, PhD;
Anthony Lubniewski, MD;
Nicole Anderson, MD;
Henry Edelhauser, PhD;
Glenn Holley, BS;
Ara Tess, RN;
Hozefa Divan, MSPH;
Mark Helmy;
Matthew Arduino, DrPH;
William R. Jarvis, MD;
for the Toxic Endothelial Cell Destructioin Syndrome Investigative Team
Arch Ophthalmol. 2000;118:1167-1176.
Background Toxic endothelial cell destruction (TECD) syndrome after intraocular ophthalmic surgery is rare and can result from exposure to a variety of toxins. During January 8 to 14, 1998, 6 patients developed TECD with corneal edema associated with unreactive or dilated pupils at Hospital A.
Methods A case patient was any Hospital A patient with TECD within 24 hours after surgery during January 5 to 14, 1998 (epidemic period). A control was any hospital A ophthalmic surgery patient without TECD during the epidemic period. The medical records of hospital A ophthalmology surgery patients during the pre-epidemic (ie, September 1, 1997-January 4, 1998) and epidemic periods were reviewed. Inductively coupled plasma atomic emission spectrometry was used to detect trace inorganic elements on sterilized surgical instruments. Cannulated surgical instruments and laboratory rinsates were perfused directly to the corneal endothelium of isolated rabbit and human corneas. Corneal endothelial ultrastructure and swelling were assessed.
Results The rate of TECD at hospital A was higher during the epidemic than pre-epidemic period (6/12 vs 0/118, P<.001). The only change during the periods was the introduction, on November 5, 1997, of a new sterilization method, AbTox Plazlyte, for sterilization of ophthalmic surgery instruments. Findings from spectrometry revealed that copper and zinc residues were higher in instruments sterilized with Plazlyte than in those sterilized with ethylene oxide (median copper value, 7.64 mg/L vs 0.14 mg/L, respectively, P = .02; median zinc value, 5.90 mg/L vs 1.35 mg/L, respectively, P = .2). Corneal endothelial perfusion of Plazlyte sterilizedinstrument rinsates or laboratory solution with copper and zinc produced irreversible damage, similar to toxic corneal endothelial destruction, to rabbit and human corneas.
Conclusion A new sterilization method degraded brass to copper and zinc on cannulated surgical instruments resulting in TECD of the cornea.
From the Hospital Infections Program, National Center for Infectious Diseases (Drs Duffy, Arduino, and Jarvis), Division of Surveillance, Hazard Evaluations and Field Studies, National Institute for Occupational Safety and Health (Ms Brown), Division of Environmental Health Laboratory Sciences, National Center for Environmental Health (Dr Caldwell), Centers for Disease Control and Prevention, Public Health Service, US Department of Health and Human Services, Atlanta, Ga; Department of Ophthalmology and Visual Sciences, Washington University, and the Ophthalmology Section, Surgical Service, St Louis Veteran Affairs Medical Center, St Louis, Mo (Dr Lubniewski); Department of Ophthalmology, Emory University School of Medicine, Atlanta (Drs Anderson and Edelhauser, Messrs Holley and Helmy); John Cochran Division, Veteran Affairs Medical Center, St Louis (Ms Tess); and the Department of Epidemiology, Rollins School of Public Health, Emory University (Mr Divan). Dr Duffy and Ms Brown are officers in the CDC's Epidemic Intelligence Service Program.
RELATED ARTICLE
Archives of Ophthalmology Reader's Choice: Continuing Medical Education
Arch Ophthalmol. 2000;118(9):1310-1311.
FULL TEXT
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
The Balance between Corneal Transparency and Edema The Proctor Lecture
Edelhauser
IOVS 2006;47:1755-1767.
FULL TEXT
Effects of Enzymatic Sterilization Detergents on the Corneal Endothelium
Parikh et al.
Arch Ophthalmol 2002;120:165-172.
ABSTRACT
| FULL TEXT
|