You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 117 No. 7, July 1999 TABLE OF CONTENTS
  Archives
  •  Online Features
  Clinical Sciences
 This Article
 •Full text
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citation map
 •Citing articles on HighWire
 •Citing articles on ISI (49)
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Radiologic Imaging
 •Alert me on articles by topic
 Social Bookmarking
  Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit Add to Technorati
What's this?

Corneal Thickness Measurements With the Topcon SP-2000P Specular Microscope and an Ultrasound Pachymeter

Renee Bovelle, MD; Stephen C. Kaufman, MD, PhD; Hilary W. Thompson, PhD; Hikaru Hamano, MD

Arch Ophthalmol. 1999;117:868-870.

Objective  To compare the reproducibility of measurements obtained with a new pachymetry instrument, the Topcon specular microscope (Topcon SP-2000P; Topcon America Corp, Paramus, NJ), with those obtained by ultrasound pachymetry.

Methods  Corneal thickness was measured in 40 eyes of 40 patients 3 times each with the Topcon SP-2000P and an ultrasound pachymeter (DGH 500; DGH Technology Inc, Exton, Pa) by 2 separate investigators. Comparisons included average thickness as measured by each instrument, average thickness for each instrument as measured by each investigator, and differences in thickness due to corneal abnormalities.

Results  Mean corneal thickness measured by the Topcon instrument was significantly less (32 µm; P<.001) than the mean value obtained with the ultrasound pachymeter. Similarly, mean values obtained with the 2 instruments by the 2 investigators were significantly different (P<.001 and .008 for investigators 1 and 2, respectively), with the Topcon value less than the ultrasound value in both cases. Both instruments detected abnormalities in corneal thickness equally well. However, the measurements obtained with the Topcon instrument by the 2 investigators were more consistent (no significant difference [P=.32]) than those obtained with the ultrasound unit (difference was significant [P=.02]).

Conclusions  The new noncontact Topcon specular microscope provides measurements of corneal thickness that are somewhat less than those of ultrasound pachymetry, but that seem to be more consistent from one operator to another, possibly as a result of the elimination of observer bias induced by probe placement required by the ultrasound unit. This consistency may be important in the comparison of measurements by different operators over time in patients being followed up after refractive surgery or other therapeutic interventions.


From the LSU Eye Center, Louisiana State University Medical Center School of Medicine, New Orleans.



Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati     What's this?

THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES

Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis
Kawana et al.
Br. J. Ophthalmol. 2004;88:466-468.
ABSTRACT | FULL TEXT  





HOME | CURRENT ISSUE | PAST ISSUES | TOPIC COLLECTIONS | CME | SUBMIT | SUBSCRIBE | HELP
CONDITIONS OF USE | PRIVACY POLICY | CONTACT US | SITE MAP
 
© 1999 American Medical Association. All Rights Reserved.