Biodegradable Mechanical Retinal FixationA Pilot Study
Karl R. Olsen, MD;
Jean-Marie Parel, Ing ETS-G;
William Lee;
Eleut Hernandez;
Steven J. Fliesler, PhD
Arch Ophthalmol. 1989;107(5):735-741.
Abstract
 |  |
Mechanical retinal fixation is a useful tool in the armamentarium of vitreoretinal surgeons. Unfortunately, the use of metallic retinal tacks on a permanent basic mandates a retained intraocular foreign body that poses some potential for longrange toxic effects. Herein, we report the results of our initial experience with a temporary, biodegradable mechanical retinal fixation device ("biopin") in rabbits. When compared with metallic tacks, the biopins appeared to be well tolerated intraocularly, with minimal damage detectable at the light microscopic level. Disinsertion of the biopins occurred in 46% of the eyes within two weeks and in all eyes by four weeks. Initial signs of degradation of the biopins were observed at six weeks (median) in the lensectomized/vitrectomized eyes and at 12 weeks (median) in the nonvitrectomized eyes. Differences were found in the occurrence of retinal detachment between animals that underwent lensectomy/vitrectomy with insertion of two metallic tacks and animals that underwent insertion of one biopin without vitrectomy. The biopin has a potential as a vehicle for sustained release of pharmacologic agents to inhibit directly the development of vitreoretinal proliferation, thereby retarding or preventing subsequent retinal traction and detachment.
Author Affiliations
From the Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami School of Medicine (Drs Olsen and Fliesler and Messrs Parel and Hernandez), and the Veterans Administration Medical Center, Miami (Dr Olsen and Mr Lee). Dr Fliesler is now with the Bethesda Eye Institute, St Louis University School of Medicine.
Footnotes
Accepted for publication Nov 9, 1988.
Presented in part at the American Retina Society Meeting, Colorado Springs, Colo, Sept 19, 1986; at the 25th Anniversary of the Bascom Palmer Eye Institute Scientific Symposium, Miami Beach, Fla, Feb 19, 1987; and at the Annual Meeting of the Association for Research in Vision and Ophthalmology, Sarasota, Fla, May 7, 1987.
Reprint requests to Bascom Palmer Eye Institute, PO Box 016880, Miami, FL 33101 (Dr Olsen).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Scleral Plug of Biodegradable Polymers for Controlled Drug Release in the Vitreous
Hashizoe et al.
Arch Ophthalmol 1994;112:1380-1384.
ABSTRACT
|