Comparison of the Tono-Pen to the Goldmann applanation tonometer
R. E. Frenkel, Y. J. Hong and D. H. Shin
Kresge Eye Institute, Wayne State University School of Medicine, Detroit, MI 48201.
Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry
in 142 eyes of 71 patients without corneal problems. Overall we found 63%
of the Tono-Pen readings to be within +/- 2 mm Hg of the Goldmann
applanation tonometry readings and 77% to be within +/- 3 mm Hg, while 4%
were within at least +/- 7 mm Hg difference. The correlation coefficient
between the readings of the two instruments was .84. The mean value of the
paired differences was 0.81 +/- 3.09 mm Hg overall, -1.78 +/- 1.86 mm Hg in
the 4 to 10 mm Hg interval, 0.07 +/- 2.97 mm Hg in the 11 to 20 mm Hg
interval, 1.27 +/- 2.62 mm Hg in the 21 to 30 mm Hg interval, and 4.15 +/-
2.28 mm Hg in the 31 to 45 mm Hg interval. We conclude that the Tono-Pen
measures intraocular pressure in a manner that corresponds well to the
Goldmann tonometer in the 11 to 20 mm Hg interval, and fairly well in the 4
to 10 mm Hg and 21 to 30 mm Hg intervals. It lacks good correspondence in
the 31 to 45 mm Hg interval.
Reproducibility and Clinical Evaluation of Rebound Tonometry
Martinez-de-la-Casa et al.
IOVS 2005;46:4578-4580.
ABSTRACT
| FULL TEXT
A comparison of four methods of tonometry: method agreement and interobserver variability
Tonnu et al.
Br. J. Ophthalmol. 2005;89:847-850.
ABSTRACT
| FULL TEXT
Pressure Phosphene Self-Tonometry: A Comparison with Goldmann Tonometry in Glaucoma Patients
Lam et al.
IOVS 2004;45:3131-3136.
ABSTRACT
| FULL TEXT
Dynamic Contour Tonometry: A Comparative Study on Human Cadaver Eyes
Kniestedt et al.
Arch Ophthalmol 2004;122:1287-1293.
ABSTRACT
| FULL TEXT
Clinical Comparison of the Proview Eye Pressure Monitor With the Goldmann Applanation Tonometer and the TonoPen
Li et al.
Arch Ophthalmol 2004;122:1117-1121.
ABSTRACT
| FULL TEXT
Effect of Corneal Thickness on Intraocular Pressure Measurements with the Pneumotonometer, Goldmann Applanation Tonometer, and Tono-Pen
Bhan et al.
IOVS 2002;43:1389-1392.
ABSTRACT
| FULL TEXT
Tono-Pen determination of intraocular pressure in patients with band keratopathy or glued cornea
Azuara-Blanco et al.
Br. J. Ophthalmol. 1998;82:634-636.
ABSTRACT
| FULL TEXT
Measurement of intraocular pressure after epikeratophakia
Wright and Grajewski
Br. J. Ophthalmol. 1997;81:448-451.
ABSTRACT
| FULL TEXT